Summary of Contents for Bioneer AllInOneCycler

  • Page 2 AllInOneCycler Thermal Block Bioneer Corporation 8-11, Munpyeongseo-ro, Daedeok-gu, Daejeon, 34302, Republic of Korea Tel: +82-42-930-8777 Fax: +82-42-930-8688 Email: sales@bioneer.com www.bioneer.com BQ-042-101-01 Revision : 0(2012-12-07) www.bioneer.com...
  • Page 3 AllInOneCycler Thermal Block AllInOneCycler ™ BQ-042-101-01 Revision : 0(2012-12-07) www.bioneer.com...
  • Page 4: Table Of Contents

    AllInOneCycler Thermal Block Contents Getting Started ....................... 1 AllInOneCycler Symbols on the ™..................3 III. Safety Warnings and Precautions ..................5 1. User and experimental precautions ................. 5 2. Precautions regarding the electrical environment ............5 3. Precautions regarding the operation environment ............6 4.
  • Page 5: Getting Started

    Bioneer Corporation.  Bioneer Corporation reserves the right to alter, modify and otherwise make changes to the instrument and manuals without prior notice. ...
  • Page 6 AllInOneCycler Thermal Block AllInOneCycler About the ™ AllInOneCycler Thermal Block is a multifunctional, peltier-based thermal block, equipped with a AllInOneCycler temperature control system. Thermal Block has various functions including temperature increment, time increment and ramp rate control. This system is compact in size and light weight for easy storage and moving.
  • Page 7: Symbols On The Allinonecycler

    AllInOneCycler Thermal Block AllInOneCycler II. Symbols on the ™ AllInOneCycler Symbols on the ™ user manual AllInOneCycler The following table describes the symbols that may be displayed on user manual. Conformité Européenne Mark In vitro diagnostic medical device Temperature limitation...
  • Page 8 ™ as unsorted municipal waste. Follow local municipal waste ordinances for proper disposal provisions to reduce the environmental impact of waste electrical and electronic Instrument. European Union customers: Call your local Europe office for Bioneer instruments pick-up and recycling. BQ-042-101-01 Revision : 0(2012-12-07) www.bioneer.com...
  • Page 9: Safety Warnings And Precautions

    III. Safety Warnings and Precautions The warnings and precautions stated below are for the correct and safe operation of the instrument. Please heed all information for your safety. Bioneer Corporation does not assume responsibility for non-compliance with the safety warnings and precautions stated below.
  • Page 10: Precautions Regarding The Operation Environment

    AllInOneCycler Thermal Block shock or fire. (2) Do not operate multiple instruments out of a single wall outlet. The load may cause overheating and may lead to fire. (3) When plugging or unplugging the power cord from a wall outlet, make sure your hands are completely dry.
  • Page 11: Precautions And Warnings Regarding Instrument Installation

    AllInOneCycler Thermal Block 4. Precautions and warnings regarding instrument installation (1) This is a precision instrument. Do not install in a location exposed to direct sunlight. (2) Install the instrument on a flat, solid surface that is flat and does not move.
  • Page 12: Precautions Regarding Moving And Lifting Instrument

    AllInOneCycler Thermal Block (1) This product must only be used for nucleic acid amplification. Do not use the instrument for any use other than explicitly stated in the User Guide. (2) You must use the supplied accessories and tubes. (3) Do not modify or delete instrument-related information installed within the instrument.
  • Page 13 AllInOneCycler Thermal Block (3) Always determine what chemicals have been used in the instrument before changing reagents or instrument components. Wear appropriate eyewear, protective clothing, and gloves when working on the instrument. (4) Read and understand the Material Safety Data Sheet (MSDS) provided by the chemical manufacturer before you store, handle, or work with any chemicals or hazardous materials.
  • Page 14: System Components And Specifications

    AllInOneCycler Thermal Block IV. System Components and Specifications 1. System Components AllInOneCycler Figure 1. System components of the BQ-042-101-01 Revision : 0(2012-12-07) www.bioneer.com...
  • Page 15 AllInOneCycler Thermal Block AllInOneCycler Table 1. System components of the Part Name Cat. No. Qty. AllInOneCycler PCR system AllInOneCycler (Basic) 96 well PCR system A-2041-1N 1 ea (Optional) AllInOneCycler 384 well PCR system A-2041-2N AllInOneCycler (Optional) Slide PCR system A-2041-3N...
  • Page 16: Specifications

    96 well thermal block (Cat. No. A-2041-1-1), and other formats of block are optional. Bioneer provides the interchangeable thermal blocks made of 2 different metals (normal and novel alloy) and classified to 3 sample-well types (96-well, 384-well and slide PCR).
  • Page 17: System Views

    AllInOneCycler Thermal Block 3. System Views Figure 2. Front view of the AllInOneCycler Figure 3. Back view of the AllInOneCycler BQ-042-101-01 Revision : 0(2012-12-07) www.bioneer.com...
  • Page 18: Installing The Allinonecycler Tm

     Scissors, pocket knife, or box cutter AllInOneCycler 3. Unpacking the Save the packing materials and box in case you need to ship the instrument to Bioneer for service. (1) To unpack the AllInOneCycler AllInOneCycler Figure 4. Package of the Cut the straps securing the instrument box.
  • Page 19: Setting Up The Allinonecycler Tm

    AllInOneCycler If the is damaged, note the location and appearance of the damage, and then contact Bioneer Technical Support or your service representative. AllInOneCycler (2) Move the to the desired installation site. Follow these guidelines for lifting and moving: ...
  • Page 20: Power On

    AllInOneCycler Thermal Block 5. Power ON AllInOneCycler (1) Confirm that the hot-top cover of the is closed. AllInOneCycler (2) Press the power button on the back left side of the AllInOneCycler Figure 5. Power button of the (3) When the initialization of the device is completed, ‘Home’ menu will be displayed on screen.
  • Page 21: Operating The Allinonecycler Tm

    AllInOneCycler Thermal Block AllInOneCycler VI. Operating the 1. Loading Samples into the Instrument  This section describes how to load samples in tubes or plates.  See the appropriate section for your instrument.  There is no need to use mineral oil or glycerin in the block wells or as a vapor barrier over the PCR reaction mixture in the tubes.
  • Page 22: Home Menu

    AllInOneCycler Thermal Block Close the hot-top cover. 2. Home menu  The user can create a new protocol, select the saved protocols, change the setup, and confirm the information of the instrument selectively on this ‘Home’ menu. Table 2. Function icons of Home menu...
  • Page 23: Editing Protocols

    AllInOneCycler Thermal Block 3. Editing Protocols (1) Creating a new protocol 1) Press ‘New Protocol’ icon of ‘Home’ menu.  Press ‘New Protocol’ icon, and then ‘Edit Protocol’ screen will appear.  In this screen, the user can select a protocol among template protocols for new experiment.
  • Page 24 AllInOneCycler Thermal Block List Description The numbers to distinguish the order of the steps. 3. Step order The protocol will be sequentially processed. Display the temperature of each step. 4. Temperature By pressing ‘the temperature’ icon, the user can change temperature values.
  • Page 25 AllInOneCycler Thermal Block 2) Select a protocol among template protocols  We provide 8 protocol template protocols as described below. The user can select one of these for their intended purpose. Table 4. Template protocol list Protocol name Standard 3 step...
  • Page 26 AllInOneCycler Thermal Block (2) Editing protocols in ‘Edit Protocol’ screen 1) Modifying the temperature, time and cycles of the existing steps. Select the values of the temperature, the time and the repeats on the procedure graph. b) ‘Step Options’ window will appear. There are 4 types of ‘Step Options’ described in Table 5.
  • Page 27 AllInOneCycler Thermal Block  Temperature range : 4 °C ~ 99.9 °C  Time range : 1 seconds ~ 5 hours 59 minutes 59 seconds ※ Set the incubate options additionally. Figure 10. Incubate options of Incubate step option (Lower left) ...
  • Page 28 AllInOneCycler Thermal Block Table 6. How to set incubate options. Option Description  This menu is to set the gradual time increase and decrease of the selected step with each cycle. Touch and change the values between 0 and 59.
  • Page 29 AllInOneCycler Thermal Block ② Goto step Figure 11. Goto step option window of existing step  Touch the ‘Goto’ icon or select number of repeats, and ‘Goto Step Options’ window will appear.  Change the 'step number' to start the repeat interval. The ‘Step Number’ must be smaller value than the step number in the ‘Goto’...
  • Page 30 AllInOneCycler Thermal Block  Store temperature : 4 °C ~ 12 °C  Store step is provided in the last part of each protocol for proper storage of samples. 2) Adding new steps. Touch the step to add a new step on their right. If a specific step is not selected, a new step will be added in front of the first step.
  • Page 31 AllInOneCycler Thermal Block 3) Deleting steps Touch the step(s) to remove. Figure 14. Selecting a step that needs to remove b) Press ‘Delete’ icon of the bottom menu. 4) Opening the previously saved protocol Press ‘Open’ icon. b) ‘Open protocol’ screen will appear.
  • Page 32 AllInOneCycler Thermal Block Figure 15. ‘Save Protocol’ window Select an existing folder in ‘Folder selection’ window. Figure 16. Selecting the existing folder to save the edited protocol d) Or if it is necessary to create a new folder, touch the check box at the left of ‘New Folder’...
  • Page 33 AllInOneCycler Thermal Block Figure 17. Creating a new folder Assign a name to the new protocol. Figure 18. Assigning a name of the edited protocol If a folder is not selected, the protocol will be saved in the "Default" folder.
  • Page 34 The user cannot create new folder directly from the USB drive in ‘Save Protocol’ window.  If there are folders in ‘USB:\AllInOneCycler\Protocols’, the user can find the folders displayed as ‘USB – folder name’ and select one to save protocols in ‘protocol selection’...
  • Page 35 AllInOneCycler Thermal Block Figure 21. Save the protocol in USB drive Press ‘OK’ button to finish. BQ-042-101-01 Revision : 0(2012-12-07) www.bioneer.com...
  • Page 36 AllInOneCycler Thermal Block (3) Opening the existing protocols 1) ‘Open Protocol’ menu Press ‘Open Protocol’ icon at ‘Home’ menu, ‘Open protocol’ screen will appear. AllInOneCycler Figure 22. Open Protocol screen of the b) Find the previously saved folders in the ‘Folders’ window on the upper left side of the screen by pressing the up and down arrows.
  • Page 37 AllInOneCycler Thermal Block Table 7. Function icons of Open Protocol screen Function icon Description Return to ‘Home’ menu. Edit the selected protocol. When pressing this icon, ‘Edit protocol’ screen will be displayed. And the user can modify the selected protocol in the same way as in the ‘New Protocol’...
  • Page 38 AllInOneCycler Thermal Block 2) Open protocols saved in USB drive  The user can open or run protocols saved in USB drive. AllInOneCycler Insert USB drive to the USB ports in front or rear (lower) of (See figure 2 and 3 at p. 13).
  • Page 39 AllInOneCycler Thermal Block 3) Editing the existing protocol. Press ‘Edit’ icon in ‘Open protocol’ menu, and the selected protocol will be displayed in ‘Edit Protocol’ screen. b) The name of current protocol is displayed on the ‘protocol name’ on the center of the top screen.
  • Page 40: Running The Pcr Protocols

    AllInOneCycler Thermal Block 4. Running the PCR protocols  Press ‘Run’ icon on ‘Edit Protocol’ or ‘Open Protocol’ screen, and then the ‘Run’ screen is displayed as shown below. Figure 25. ‘Run’ screen Table 8. The components of ‘Run’ screen...
  • Page 41 AllInOneCycler Thermal Block List Description Show the running time. (H:M:S) When error occurs on the peltier, error messages will be displayed instead of time Return to ‘Home’ menu. ‘Pause’ is to stop temporarily. 6. Function icons ‘Resume’ is to continue ‘Run’ again from the ‘Pause’...
  • Page 42: Removing The Tubes Or Plate From The Allinonecycler Tm

    AllInOneCycler Thermal Block Figure 27. Edit Protocol screen during PCR running  If the lid is open, the current protocol will be paused. Use caution because the lid is still hot. AllInOneCycler 5. Removing the tubes or plate from the ...
  • Page 43: Setting Menu

    Software update AllInOneCycler Before updating software of , download the recent version from the Bioneer website (www.bioneer.com). Once most current software has been downloaded,create new folder named ‘USB:\AllInOneCycler_Update’ in USB, and then copy the software to the folder. AllInOneCycler b) Insert USB drive to the USB ports in front or rear (lower) of (See figure 2 and 3 at p.
  • Page 44 AllInOneCycler Thermal Block Figure 29. Software update Press ‘S/W Update’ button, and the software will be installed automaticaly, and then the instrument will restart. Setting the instrument name  Touch ‘Instrument Name’ window, and the keypad will appear on the bottom of the screen.
  • Page 45 AllInOneCycler Thermal Block (2) Network  Set the network settings including IP address for the remote control. Figure 31. ‘Network setting’ menu  Touch the values that need to change, and the number pad will appear on the center of the screen.
  • Page 46 AllInOneCycler Thermal Block (3) Data  The user can confirm and delete the protocol files, and also copy them to USB drive. Protocol confirmation Press ‘Data’ button, and the file manager will display on the right part of screen. Figure 33. Data manager screen b) When selecting ‘Protocol’...
  • Page 47 AllInOneCycler Thermal Block To confirm protocol, press the name of the files twice. Details of the selected file will be displayed on screen. Figure 34. Protocol graph view screen Copying protocol files to USB drive Insert USB drive to the USB ports in front or rear (lower) of...
  • Page 48 AllInOneCycler Thermal Block After selecting folder or file by pressing their name, press ‘Save’ icon in the bottom menu. If only selecting folder, all files in the folder will be saved. Figure 36. Selecting a folder and protocol file d) The success message ‘Completed’ will be displayed.
  • Page 49 AllInOneCycler Thermal Block In USB drive, the new folder named ‘AllInOneCycler’ will be created. The protocol files are saved to ‘Protocols’ folder of ‘AllInOneCycler’. If the user selects a folder and presses the ‘Save’ icon of the bottom menu, the folder will be copied in the ‘Protocols’...
  • Page 50 AllInOneCycler Thermal Block (4) Option  This menu is about resetting to factory mode (only for the engineer).  A mouse can be used for this mode. AllInOneCycler Connect USB mouse controller to the front or back USB port (lower) on (See figure 2 and 3 at p.
  • Page 51: About Menu

    AllInOneCycler Thermal Block (5) Functions  In ‘Setting’ menu, there are four functions that are described in the following table. Table 9. Function icons of ‘Setting’ menu Function icon Description Calibrate the temperature control setting of the instrument. (Only for the engineer) Save the current settings for each sub-menu and save protocol files to USB drive.
  • Page 52: Network Operating Allinonecycler Tm

    AllInOneCycler Thermal Block AllInOneCycler 8. Network Operating (1) Composition  AllInOneCycler It is possible to operate multiple instruments through TCP/IP network.  The optional PC is necessary in order to control network operation.  The PC requires ‘Server.exe’ program and ‘Manager.exe’ program for controlling.
  • Page 53 AllInOneCycler Thermal Block AllInOneCycler d) Turn on the instrument(s), ‘Poplar.exe’ operating will automatically start. When turning on the instrument, the network between the instrument and PC should be connected. AllInOneCycler Figure 44. Connection status between PC and Run ‘manager.exe’ in PC, and the connection state will be displayed as shown below.
  • Page 54 AllInOneCycler Thermal Block Figure 45. Connection status when 'manager.exe' program is running in PC When ‘manager.exe’ is running, the currently connected instrument is displayed as shown below. Figure 46. 'manager.exe' window shows the currently connected instrument BQ-042-101-01 Revision : 0(2012-12-07)
  • Page 55 AllInOneCycler Thermal Block (3) Execution of running Select a protocol Click the ‘Batch Run’ icon in the ribbon menu. Figure 47. The ribbon menu of 'manager.exe' program b) The ‘Batch Run’ window will appear, and then the user can select and run the protocol saved in local (PC) or the connected instrument.
  • Page 56 AllInOneCycler Thermal Block Select the protocol by using the protocol file tree in the left top side of the screen, from the file explorer. In this figure, the protocol named ‘Demo2’ is selected. d) When a protocol is selected the corresponding protocol graph will be displayed in the right top side of the screen.
  • Page 57 AllInOneCycler Thermal Block Figure 50. Confirmation message of starting the selected protocol in the selected instrument Stop When the instrument is running, ‘Stop’ button is activated in the lower right of the screen. In addition, the user can pause or resume the current running PCR by clicking the ‘Pause’...
  • Page 58 AllInOneCycler Thermal Block b) If ‘Stop’ button is clicked, the stop signal appears in the screen of the running instrument, and then the instrument will stop. AllInOneCycler Figure 52. Stop screen of the controlled by PC If ‘OK’ button is clicked, the instrument will stop. And then, the main menu will be displayed on the screen.
  • Page 59: Troubleshooting

    AllInOneCycler Thermal Block VII. Troubleshooting Error Solution The Power is not on. Make sure the power connector is connected. Check if the instrument is connected to the power cord. Check if the instrument's power switch is turned off. If everything above has been followed but the power is not on, contact Service.
  • Page 60 AllInOneCycler Thermal Block Error Solution It stops while operating. Check if power-supply to the instrument is on. Check if the Power switch has been pressed off. Check if you have pressed ‘Pause” by mistake. Request for Service. The System is running but there Check if wrong protocol has been used.
  • Page 61 AllInOneCycler Thermal Block buttons pictures Erase and re-install the software. displayed within the software. BQ-042-101-01 Revision : 0(2012-12-07) www.bioneer.com...
  • Page 62: User Maintenance

    AllInOneCycler Thermal Block VIII. User Maintenance 1. Cleaning the main body Note  Power off the AllInOneCycler Thermal Block, and then remove the power cord.  Allow the instrument to cool until the hot-top cover and bath block reach room temperature.
  • Page 63: Cleaning The Hot-Top Cover

    AllInOneCycler Thermal Block 2. Cleaning the Hot-top Cover  Clean the hot-top cover once a month or as needed. During instrument operation, the temperature of the hot-top cover can be as high as 110°C, and the temperature of the bath block can be as high as 100°C. Before cleaning instrument, ensure that the hot-top cover and bath block reach room temperature.
  • Page 64 AllInOneCycler Thermal Block Procedure (1) Turn off the instrument and disconnect the power cord from the back of the AllInOneCycler ELECTRICAL SHOCK HAZARD Severe electrical shock, which could cause physical injury or death, can result from working on an instrument when the high voltage power supply is operating. To avoid electrical shock, disconnect the power supply to the instrument, unplug the power cord, and wait at least 1 minute before working on the instrument.
  • Page 65: Returning An Allinonecycler Tm For Service

    Bioneer for service. (2) Complete instrument decontamination. (3) Fax or e-mail(sales@bioneer.com) the service request form to the customer center. (4) Pack the instrument in the provided packaging, without any accessories or power cords. Include service request form in the box.
  • Page 66: Warranty

    (1) year from the date you received your product. Bioneer will either (1) repair the product at no charge if a hardware defect is found or (2) exchange the product if the same hardware defect arises more than three times during the limited warranty period.
  • Page 67 Issue 1 Issue 2 Service Issue Issue 3 Issue 4 Address Company Customer Name Information Contact Phone: Fax: E - mail  Please send service request to Fax(042-930-8688) or E-mail(sales@bioneer.com).  You can download this service request document in Bioneer website(www.bioneer.com).