Bioneer ExiPrep 16 Dx User Manual page 45

Fully automated nucleic acid extraction system
Hide thumbs Also See for ExiPrep 16 Dx:
Table of Contents

Advertisement

Error
The base plate does not slide in
completely.
The system is not working even if
the door is closed.
The base plate does not slide out
completely.
The accessories are not being
able insert.
During the operation, the syringe
block does not pin down the tips.
Syringe
Block
is
equipped with the tips, but it
doesn't move.
A4(210X297)
1.
Check if there are any obstacles or residues that may interrupt
moving the base plate.
2.
Check if all the accessories are inserted correctly.
3.
Request for After Service.
1.
Check if base plate is located in the correct.
2.
Check if the pair of magnet attached to the door is correctly
attached.
3.
Check if the switch installed at the end of the base plate slide
rail is damaged when base plate is pulled out.
4.
Request for After Service.
1.
Check inside of the system if there are any obstacles or
residues that may interrupt moving the base plate.
2.
Check the front of the system if there are any obstacles or
residues that may interrupt moving the base plate.
3.
Request for After Service.
1.
Check if all the accessories are located in the right position.
2.
Check if there are any residues in each accessories and the
rack.
3.
Check if any accessories and the rack's lock pins are bent or
damaged.
4.
Request for After Service.
1.
Check if the door is completely closed.
2.
Check if the tip and the tip rack is inserted correctly.
3.
While inserting the tips, check if any residues or accessories
are stuck.
4.
Check if 'Stop' or 'Pause' button is pressed.
5.
Check if the provided tips are inserted.
6.
Check if the tip is bent or damaged.
7.
Request for After Service.
correctly
1.
Make sure the front door is closed.
2.
When the block is being moved, check if any residues or
accessories are stuck.
3.
Check if the cartridge is inserted correctly.
4.
Check if 'Stop' or 'Pause' button is pressed.
5.
Request for After Service.
Fully Automated Nucleic Acid Extraction System
Solution
-43-
TM
ExiPrep
16 Dx