※ Setup process according to the CHECK LIST
.
A4(210X297)
Place the Buffer Cartridge ② on the heating block of the base plate.
1.
If Buffer Cartridge ② is not properly placed on the heating block, it
may cause an experiment failure or an instrument malfunction.
2. Place the Buffer Cartridge
Place the Buffer Cartridge ① by putting the cartridge to the left side
of the base plate and press right-hand side of the Cartridge to press
the Cartridge to secure.
3. Place the Sample tube rack on the base plate.
Be careful that the Sample tube rack's front and rear sides are
properly placed.
4. Place the Waste tray in between the Sample tube rack and the Buffer
Cartridge ②.
5.
Place the Elution tube rack on the base plate.
Check the Protection Cap is properly secured on the elution tube
rack.
WARNING!
Be careful with the Aluminum sheet of the Protection Cap.
6. Place the Disposable tip rack on the base plate.
7. Push the Base plate all the way inside the instrument and close the
door.
-17-
ExiPrep
Fully Automated Nucleic Acid Extraction System
on the base plate.
TM
16 Dx