V. DNA/ RNA Extraction

1. Contamination Protection Accessory Installation

A4(210X297)
Fully Automated Nucleic Acid Extraction System
1.
From the 'Menu' screen, click 'MISC SET'.
Pull out the Syringe block all the way front (outside)
Contamination Shield and Tip Protector.
NOTE
"Load" button of ExiStation
with "MISC SET" button.
CAUTION
In order to use protection accessory, this must be installed on
the Base plate accessory.
2.
Place the Tip Protector on top of the Syringe block, and insert it
into two holding shafts.
Make sure front and rear are properly inserted.
Do NOT FORCE to insert the Tip Protector. It may damage.
3.
Place the Contamination Shield on the lower-right side of the
Syringe block.
Contamination Shield has a magnet which means if you
place on the upside-down right, it will stick to the holding bar.
4.
From the 'Menu' screen, click 'MISC SET'.
The base plate moves back to the initialization position
(inside the instrument).
NOTE
"Load" button of ExiStation S/W perform the same function
with "MISC SET" button.
-11-
TM
ExiPrep
TM
S/W perform the same function
16 Dx