3. RUN( Not used PC) (Continued)
A4(210X297)
Fully Automated Nucleic Acid Extraction System
13. After the extraction run is complete, pull out the Base Plate and
remove the Elution tubes, Buffer Cartridges and all racks from
the Base Plate. After removing all accessories, push the Base
Plate back in completely and close the door.
14. You are given three options at this point:
Still remains same work: Repeat the current protocol.
Do other work: Perform an extraction run using a different
protocol for another nucleic acid and sample source type.
Finish: Finish and exit.
15. If the automatic UV-sterilization option is enabled, a popup
prompt will appear warning you not to open the door as UV
sterilization is in progress. Details can be found in
16. Press the 'START' button to initiate sterilization
Select 'SKIP' if you wish to pass sterilization.
17. The sterilization process takes 5 minutes. Progress can be
checked through the progress bar.
18. Remove the contamination protection accessory (Tip protector,
Contamination shield). Detail can be found in
19. Press the 'Store On' button to turn the Cooling Block and
cooling fan off.
-16-