Figure 3-5. Manually Pausing Alarms - Covidien Puritan Bennett 560 User Manual

Ventilator
Table of Contents

Advertisement

Alarms and Troubleshooting
Warning
Alarm volume should be adjusted with respect to the ventilator's operating environment and so
that the patient's caretakers can hear the alarms. The audible alarm vents located at the front of
the device should never be obstructed. The alarm can be paused with the Alarm Pause function
by pressing the ALARM CONTROL key twice once the alarm has been declared.
When an alarm condition is triggered, or there is evidence of a patient-ventilator fault or
problem, examine the patient first before examining the ventilator.
Some alarms are not automatically cancelled when the condition causing the alarm clears e.g. HIGH PRES-
SURE. Some alarms can be paused manually even if the cause(s) of their activation remain(s).
To manually pause an alarm, proceed as follows:
Press the ALARM CONTROL
The alarm is paused until the alarm condition is corrected and the condition reoccurs: the
audible portion, light indicator, and message are all halted (for the alarms which can be
paused manually).
The Alarm Paused
Alarms, and Waveforms screens. Refer to the following graphic:
When no other alarms are currently activated, the last alarm cancelled is displayed continuously in the alarm mes-
sage window in the Alarms menu, along with the date and time of its activation. The High Pressure alarm must
be manually reset. Refer to section
To manually reset the High Pressure Alarm, proceed as follows:
Press the ALARM CONTROL
The visual alarms will be reset.
3-6
key twice.
symbol is displayed at the top right of the Ventilation,

Figure 3-5. Manually Pausing Alarms

3.7, "Overview of
Alarms".
key twice.
User's Manual

Advertisement

Table of Contents
loading

Table of Contents