Specifications
Table A-10. Ventilator Range, Resolution, and Accuracy (Continued)
Ventilator Settings
PEEP
Rise time
Backup rate
Apnea time
Minimum Inspired Tidal Volume
(Min VTI)
Maximum Inspired Tidal Volume
(Max VTI)
Minimum Exhaled Tidal Volume
(Min VTE)
Maximum Exhaled Tidal Volume
(Max VTE)
Maximum Respiratory Rate
(Max Rtot)
A-6
Range, Resolution, and Accuracy
Range: OFF (0.5 mbar) to 20 mbar
Resolution: 1 mbar
Accuracy: ± (1 mbar + 10%) mbar
Default value: OFF
Depends on: PIP in P A/C and PSV modes when Relative
Pressure is set to YES
Depends on: P Support and P Control in P SIMV mode when
Relative Pressure is set to YES
Depends on: P Support in V SIMV mode when Relative
Pressure is set to YES
Range: 1-4
Resolution: 1
Default value: 2
Depends on: Insp time
Range: OFF or 4-40 bpm
Resolution: 1 bpm
Default value: 13
Depends on: Min I time
In P SIMV and V SIMV, Backup rate = Max (8, R-Rate)
Range: AUTO or 1-60 s
Resolution: 1 s
Default value: AUTO
Depends on: Backup R
In PSV, Apnea time: AUTO = 60/Backup R
In V SIMV or P SIMV, Apnea Time: AUTO = 12
In CPAP, Apnea Time: AUTO = 30
Range: 30 mL to 1990mL
Resolution: 10 mL
Default value: 300
Depends on: Max VTI
Range: 80 mL to 3000 mL
Resolution: 10 mL
Default value: 2000 mL
Depends on: Min VTI
Range: 30 mL to 1990 mL
Resolution: 10 mL
Default value: 300
Depends on: Max VTE
Range: 80 mL to 3000 mL
Resolution: 10 mL
Default value: 1000
Depends on: Min VTE
Range: 10 bpm to 70 bpm
Resolution: 1 bpm
Default value: OFF
Depends on: R-Rate