Covidien Puritan Bennett 560 User Manual page 115

Ventilator
Table of Contents

Advertisement

Table A-10. Ventilator Range, Resolution, and Accuracy (Continued)
Ventilator Settings
Pressure support
(P Support)
I:E Ratio (I:E)
I/T Ratio (I/T)
Inspiratory time (Insp Time)
Respiratory rate (R-Rate)
Inspiratory sensitivity (I Sens)
Exhalation sensitivity (E Sens)
Ramp (Flow Pattern)
User's Manual
Range, Resolution, and Accuracy
Range: OFF or 5 mbar to 55 mbar in valve configuration
Range: 6 mbar to 30 mbar in leak in valve configuration
Resolution: 1 mbar
Accuracy: ± (1 mbar + 10%) of P Support + PEEP setting
Default value: 15 mbar
Depends on PEEP when Relative Pressure is set to YES
Range: from 1:1 to 1:4
Resolution: 1/0.1 s
Accuracy: ± 50 ms or 10%, whichever is greater
Default value: 1/2
Range: 20% to 50%
Resolution: 1%
Accuracy: ± 50 ms or 10%, whichever is greater
Default value: 33%
Range: 0.3 s to 2.4 s
Resolution: 0.1 s
Accuracy: ± 50 ms or 10%, whichever is greater
Default value: 1.5 s
Depends on: R-Rate, Vt in V SIMV mode
Depends on: R-Rate in P SIMV mode
Range: 5 bpm to 60 bpm in V A/C and P A/C modes
1 bpm to 40 bpm in P SIMV and V SIMV modes
Resolution: 1 bpm
Accuracy: ± 1 bpm
Default value: 13
Depends on: Insp Time and Vt in V SIMV mode
Depends on: Insp Time in P SIMV modes
Depends on: Vt in V A/C mode
Range: 1P-5
Resolution: 1
Accuracy: NA
Default value: 2
in CPAP, I Sens is set to 2 and is not adjustable
Range: 5% to 95% of peak flow
Resolution: 5%
Accuracy: ± (4 lpm +10%) of target exhalation flow based
on E Sens within 50ms
Default value: 25%
In CPAP, E Sens is fixed at 25% and is not adjustable.
Range: Square (SQ), descending ramp (D), sinusoidal (S)
Resolution: N/A
Default value: Descending ramp (D)
In V SIMV, flow pattern is set to square and is not adjustable
Specifications
A-5