Table A-21. Electromagnetic Immunity

The ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the
ventilator should ensure that it is used in such an environment.
Immunity Test
Electrostatic discharge
(ESD)
IEC / EN 61000-4-2
Electrical fast
transient/burst
IEC / EN 61000-4-4
Surge
IEC / EN 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC / EN 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC/ EN 61000-4-8
Note: U
is the AC mains voltage prior to application of the test level.
T
A-10
IEC / EN 60601 Test Level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV lines/lines
± 2 kV lines/earth
< 5% U
T
(> 95% dip in U
for
T
0.5 cycle)
40% U
T
(60% dip in U
for 5 cycles)
T
70% U
T
(30% dip in U
for 25 cycles)
T
< 5% U
T
(> 95% dip in U
for 5 s)
T
3 A/m
Compliance Level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV lines/lines
± 2 kV lines/earth
< 5% U
T
(> 95% dip in U
for
T
0.5 cycle)
40% U
T
(60% dip in U
for 5 cycles)
T
70% U
T
(30% dip in U
for 25 cycles)
T
< 5% U
T
(> 95% dip in U
for 5 s)
T
3 A/m
Electromagnetic
Environment–Guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with synthetic
material, the relative humidity
should be at least 30%.
AC power ("mains") quality
should be that of a typical
commercial or hospital
environment.
AC power ("mains") power
quality should be that of a
typical commercial or hospital
environment.
AC power ("mains") power
quality should be that of a
typical commercial or hospital
environment. If the user of the
ventilator requires continued
operation during power mains
interruptions, it is
recommended that the
ventilator be powered from an
uninterruptible power supply or
a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.