A new message appears that directs the user to press the key again to confirm ventilation
stop. (shown in the graphic below).
A double "beep" sounds.
3. Release the VENTILATION ON/OFF
4. Press the VENTILATION ON/OFF
ventilation will continue.
Ventilation stops.
The blue LED located to the upper-right of the VENTILATION ON/OFF
item 2) illuminates to indicate ventilation is on Standby.
A prompt for a new start of ventilation is displayed (see
5.5

Turning Off the Ventilator

Warning
When the ventilator is switched back on, it will immediately begin ventilating—without the
user first having to press the VENTILATION ON/OFF
Handle the ventilator with care after use, particularly when ambient temperatures are high.
Some ventilator surfaces may be very hot, even if safety specifications are not exceeded.
Set the I/O switch to the O position to power off the ventilator.
The blue LED to the right of the VENTILATION ON/OFF
The ventilator screen switches off.
Note:
When the ventilator is completely stopped, but is still connected to the AC power source (the green
AC POWER indicator is illuminated), the internal battery continues charging.
A continuous alarm condition will be activated if the ventilator power switch is turned off while ventilation
is in progress. When the power switch is turned back on again, the ventilation will resume without having
to press the VENTILATION ON/OFF button.
User's Manual
Figure 5-10. Stopping Ventilation (2)
key.
key within 5 seconds to confirm stop, otherwise
Operating Procedures
key
Figure 5-7
on page 5-7).
key.
key turns off.
(Figure
5-8,