Operating Procedures
The parameter selected to be modified flashes.
3. Press UP
4. Press ENTER
The new parameter setting is displayed continuously.
The cursor is placed at the STOP position.
5. To manually stop trend transfer, press ENTER
If a parameter change is not confirmed by pressing ENTER
tor resets the parameter to its previous value.
Note:
The message "TRANSFER IN PROGRESS... REMAINING TIME" is displayed during the transfer time.
Other USB memory device functions are available during transfer of trends.
If the memory capacity on the USB memory device is insufficient the message "TRANSFER NOT ALLOWED -
USB CAPACITY INSUFFICIENT" is displayed and data transfer is not allowed. Delete the data on the USB
memory device before restarting data transfer.
In case of USB memory device disconnection or transfer error, the message "TRANSFER ERROR - USB
DISCONNECTION" or "TRANSFER ERROR - TECHNICAL PROBLEM" is displayed. In this case restart the
transfer process. If the problem persists contact your technical service.
5-6
or DOWN
to confirm the new parameter setting.
Table 5-2. Time taken to transfer trends data from the ventilator to a USB memory device
Amount of trends data
(in months)
3 months
6 months
9 months
12 months
to change the selected parameter's value.
.
before seven (7) seconds elapse, the ventila-
Transfer time from
ventilator to USB memory
device
Approximately 2 minutes
Approximately 4 minutes
Approximately 6 minutes
Approximately 8 minutes
User's Manual