Humidifier; Figure 4-12. Bacteria Filter; Figure 4-13. Humidifier - Covidien Puritan Bennett 560 User Manual

Ventilator
Table of Contents

Advertisement

Installation and Assembly
Connected to the FROM PATIENT
This filter protects the internal exhalation flow sensor from the gases exhaled by the patient.
Refer to
Refer to the manufacturer's instructions for more information about the use and maintenance of the bacteria
filter(s).
4.6

Humidifier

The humidifier
inserted into the patient circuit between the main outlet and the patient (refer to
Warning
During invasive ventilation (when an artificial airway bypasses the patient's upper respiratory
system), the patient's upper respiratory system cannot humidify the incoming gas. For this
reason, the use of a humidifier, to minimise drying of the patient's airways and subsequent
irritation and discomfort, must be used.
Always position a humidification device so that it is lower than both the ventilator and the
patient. Use water traps, if necessary, to limit water in the patient circuit and periodically empty
these water traps.
If a heated humidifier is used, you should always monitor the temperature of the gas delivered
to the patient. Gas delivered from the ventilator that becomes too hot may burn the patient's
airway.
Adding accessories to the ventilator breathing circuit, such as a humidifier and water trap(s),
may result in a decrease in tidal volume delivered to the patient due to the added compressible
volume of the accessory. Always assure that the patient is receiving the appropriate inspired
volume when altering the breathing circuit configuration.
When a humidification device is used, any condensation that forms in the patient circuit is collected in the
4-12
Figure
4-8, item 10.
(Figure
4-13) adds moisture (water vapour) and warms the gas in the patient circuit. It is
port:

Figure 4-12. Bacteria Filter

Figure 4-13. Humidifier

Figure 4-7
and
Figure
4-8).
User's Manual

Advertisement

Table of Contents
loading

Table of Contents