A.3

Indicators and Alarms

Table A-5. Power Indicators

Ventilation ON/OFF
Blue in standby mode
Not lit if ventilation is in
progress.

Table A-6. Alarm Indicators

High Priority
Red flashing LED
Audio Alarms
Table A-7.
Audio
Paused
60 s ± 1 s
A.4

Performance

A.4.1 Specifications

Table A-8. Performance Parameter Specifications and Tolerances

Settings
Volume
Pressure
Time
Rate
Inspiratory Sensitivity
Exhalation Sensitivity
Vt Sigh
I:E
I/T
A.5

Monitored Parameters

Table A-9. Monitored Parameter Tolerances

Ventilator Parameters
Peak Inspiratory Pressure (PIP)
Positive End Expiratory Pressure (PEEP)
Inspiratory Tidal Volume (VTI)
User's Manual
AC power
Green
Medium Priority
Yellow flashing LED
Alarm Volume
65 to 85 dBA ± 10% at 1 meter
Range
50 to 2000 ml
5 to 55 mbar
0.3 to 2.4 s
1 to 60 bpm
1 P to 5
5 to 95%
Vt x1 to Vt x 2
1:4 to 1:1
20% to 50%
Tolerances
± (2 mbar + 8%)
a
± (2 mbar + 8%)
± (10 ml + 10%VTI)*Rate
DC power
Internal Battery
Green
Flashing if the battery
charge is in progress.
Continuously lit if the
ventilator is powered by
the internal battery.
Tolerances
± (10 ml + 10%)
± (1 mbar +10%)
± 50 ms or 10%, whichever is
greater
± 1 bpm
N/A
± (4 lpm + 10%) of target
exhalation flow based on E Sens
within 50ms
± (20ml + 20%)
± 50 ms or 10%, whichever is
greater
± 50 ms or 10%, whichever is
greater

Specifications

A-3