Table 3-1. Overview of Alarms (Continued)
Alarm Message
EMPTY BATTERY
EXH VALVE
LEAKAGE
FIO
SENSOR
2
MISSING
HIGH / LOW
BATTERY TEMP*
*IF PERSISTS
RESTART/SRVC
HIGH FIO
HIGH INT TEMP
COOL VENT*
*IF PERSISTS
RESTART/SRVC
HIGH LEAKAGE
HIGH PRESSURE
User's Manual
Cause/Ventilator Response
Internal battery capacity < 10 min.
or 3%.
(battery voltage < 22.5V)
Consequence: ventilation comes to
a halt.
Abnormally high expired flow during
the inspiratory phase of three
consecutive breaths (in double-limb
setup).
Alarm activation occurs:
After three consecutive breaths.
No FIO
sensor detected and the
2
FIO
alarm is active.
2
Battery temperature out of
tolerance.
Consequence: battery charging
stops.
The level of oxygen delivered by the
2
ventilator exceeds the Max FIO
set.
Device internal ambient temperature
out of tolerance range.
The LEAK estimated by the
ventilator exceeds the Max LEAK
alarm threshold.
In V A/C or V SIMV modes, if
Inspiratory Pressure is higher than
Max PIP during three consecutive
cycles.
In PSV, CPAP, P A/C, or P SIMV
modes, if Inspiratory Pressure is
higher than (P Support or P Control
+ PEEP) + 5 mbar up to 29 mbar or
+ 10 mbar over 30 mbar during
three consecutive cycles.
In PSV or CPAP mode and P Support
is set to off, if Inspiratory Pressure is
higher than PEEP + 10 mbar during
three consecutive cycles.
Alarm activation occurs:
After three consecutive breaths.
Consequence:
Switch to exhalation phase.
Priority
level
2
or
condition
priority
indicator
must be
manually
or
reset by
pressing
the
Alarms and Troubleshooting
Audio
Paused
Avail.
HP
No
MP
Yes
HP
Yes
MP
Yes
MP
Yes
MP
Yes
MP
Yes
HP
Yes
Note:
When
portion of the
alarm
alarm may be
clears,
alarm
key.
Alarm
Paused
Avail.
No
No
Yes
Yes
No
Yes
No
No
(The visual
paused)
3-11