Table A-1. Physical Description (Excluding Accessories)

Ventilator Weight
Ventilator Dimensions
Connectors
Device airway volume
Breathing circuit volume
Adult, dual limb
Pediatric, dual limb
Adult, single limb
Pediatric, single limb
Air Inlet Filter
Inspiratory Bacteria
Filter Requirement
A.2

Electrical

Table A-2. AC Electrical Supply

Voltage
100 VAC to 240 VAC
12 VDC
30 VDC
9.9 lb. (4.5 kg)
9.25 in wide x 12.40 in deep x 6.0 in high
(235 mm wide x 315 mm deep x 154 mm high)
Inspiratory limb connector: ISO 22 mm (OD) conical
Exhalation limb connector (on exhalation block): ISO 22 mm (ID)
conical
Oxygen inlet: Female Connector with valve
2000 ml
1150 ml
670 ml
550 ml
300 ml
Dimensions: 70 mm long x 60 mm wide
Composition: Polypropylene fiber electrostatic filter material, which is
laminated onto polyurethane open-celled foam.
Efficiency: 99.999982% at 30 lpm (filtering microbes 3.3 μm)
Maximum allowable flow resistance: 4mbar at 60 lpm
Frequency
Consumption
50 Hz / 60 Hz
180 VA max
NA
8.3 A
NA
3.3 A