Configuration; Select Measurement Function: Ph/Redox Mv/Redox Mv; H /Rh/Ther - JUMO 202710 Operating Instructions Manual

Hand-held meter
Hide thumbs Also See for 202710:
Table of Contents

Advertisement

3 Configuration

Start
configuration
Select
parameter
Adjust
parameter
End
configuration
3.1
Select measurement function:
pH/redox mV/redox mV
10
Set
Press the
key for 2 seconds,
Menu
4
this calls up the configuration.
Briefly press the
4
Press the
key or the
max
2
Store
Press the
key
Quit
6
configuration is closed and the changes are saved.
Start configuration.
Select the parameter "lnP".
min
Use the
or
max
2
5
The arrow shows the selected measurement function.
rH
mV
mV
pH
H
max
2
mbar
min
5
ATC
ok
CAL
AL Logg
pH
pH: pH measurement with a pH combination electrode.
mV: Redox measurement with a redox combination electrode.
mV
As for mV. The indicated value is converted to the standard hydrogen
H
system, temperature-independent (DIN 38 404). This requires the
connection of the temperature sensor (ATC) or a manual temperature
entry.
rH
The rH value is calculated from the measured pH, redox and
temperature values. The pH and temperature values can alternatively be
entered by hand.
thEr Temperature measurement with a Pt100 temperature sensor.
The main display shows the present temperature.
After pressing the
display can show the Min or Max value, or the present measured value.
Set
key.
Menu
min
key.
5
/rH/thEr
H
key to move the arrow on the top edge of the display.
Store
key, with the
Quit
6
or
key, the subsidiary
min
max
2