As Needed Maintenance
Power Meter Window Replacement Kit
7. Insert the fused silica window over the washer. Ensure that there are no smudges or debris
on the window- clean with lens tissue and isopropyl alcohol if necessary.
Figure 23: Fused Silica Window
8. Insert a 10-mm (25/64-inch) silicon O-ring in the groove in the bottom of the knife edge cap.
Figure 24: 10mm O-ring in Groove
Page 124