Technical Data - Dentsply Sirona Xios AE Operating Instructions Manual

Sensor usb interface and sensors
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions Xios AE Sensor USB interface and sensors
67 29 433 D3780
D3780.201.01.01.02    09.2019

Technical data

3.9
AE Sensor USB interface
Electric class:
Type of protection
against electric shock:
Degree of protection
against ingress of water:
USB port:
Operating voltage:
Supply current:
Power consumption:
Maximum USB cable
length:
Dimensions L x W x H:
Weight:
Xios AE Sensors
Technology:
Physical pixel size:
Line pairs:
Measured resolution:
Theoretical resolution:
Active sensor area:
External dimensions:
Cable length:
Operating and transport conditions
Mode of operation:
Additional information:
Transport and storage
conditions:
Protection class II
Applied part type BF
Usual device (without protection against
ingress of water), corresponds to protec-
tion class IPX0
Version 2.0 and 3.0
5 V DC
Device is supplied with power via the USB
connection of the PC.
0.5 A
1.25 W
5 m
107 x 62 x 28 mm
approx. 50 g
CMOS-APS (Active Pixel Sensor)
15 µm, image acquisition in 15 µm
33.3 lp at 15 µm
28 Lp/mm
33.3 Lp/mm
Size 0 - sensor = 18 x 24 mm
Size 1 - sensor = 20 x 30 mm
Size 2 - sensor = 25.6 x 36 mm
Size 0 - sensor = 19.4 x 28.2 x 7.5 mm
Size 1 - sensor = 21.5 x 34.2 x 7.5 mm
Size 2 - sensor = 27.1 x 40.2 x 7.5 mm
max. 2.70 m
Continuous operation
The unit must not be used near flammable
anesthesia gas mixtures containing air,
oxygen or nitrogen oxide.
Temperature: -25 °C (-13 °F) to 70 °C (158
°F)
Relative humidity: 20 % to 85 %
Air pressure: 500 to 1060 hPa
3 System description
3.9 Technical data
25

Advertisement

Table of Contents
loading

Table of Contents