Installing The Adapter Plate; Preparing The Adapter Plate; Using The Proximally Located Connector - Otto Bock Harmony E2 4R152 Instructions For Use Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 16
5) After the charging process is complete, disconnect the plug from the charging receptacle and
close the charging receptacle cover.
6) Disconnect the battery charger plug from the outlet.
7) If necessary: Connect the outlet hose and attach the vacuum pump to the adapter plate.
Light Signal
LEDs 1 and 2 flash alternating yellow
(until automatic shutdown)
LEDs 1 and 2 flash alternating red (until
automatic shutdown)
LEDs 1 and 2 flash alternating green
LEDs 1 and 2 flash alternating green,
brightness dimmed
LEDs 1 and 2 illuminated green, bright­
ness dimmed
LEDs 1 and 2 illuminated turquoise,
brightness dimmed

5.2 Installing the Adapter Plate

The adapter plate for connecting the vacuum pump is installed between the prosthetic socket and
the distal components of the modular prosthesis. It is mounted together with the socket adapter.
The adapter plate has two connectors for generating a vacuum in the socket and one connector
for connecting to the vacuum pump. The adapter plate is additionally equipped with a PE sinter
filter to protect the vacuum pump against dirt, and a non-return valve that maintains the vacuum
when the pump is disconnected.
CAUTION
Using a wrong prosthetic socket
Contusions and pressure sores due to wrong socket fit
Use the vacuum pump only in combination with a prosthetic socket made with Harmony
technology.

5.2.1 Preparing the Adapter Plate

A vacuum can be generated in the socket either via the proximal or the lateral connector of the
adapter plate.
INFORMATION
Using the proximal connector of the adapter plate is recommended only when using a 5R2 lam­
ination disc or the 4A94=3 TEC plate (with TPU insert respectively). The lateral connector for
the intake hose can be recognised by the internal thread for the hose connector.

Using the proximally located connector

INFORMATION
To ensure an optimum sealed connection to the socket, the following approach is recommen­
ded:
Drill the distal socket bore with a 6 mm bit.
Apply PU adhesive inside the bore and insert the included hose piece (see fig. 12).
Remove excess adhesive and allow the adhesive to set.
Trim the hose piece flush on both ends.
20
Audible Signal
Signal tone ringing (1 x)
Signal tone ringing (3 x)
Status
Batteries almost drained, remaining
capacity ≤ 1 h
Batteries drained, device shuts
down
Charger plugged in, connection
established
Batteries are being charged
Batteries fully charged
Batteries fully charged, reduced
battery performance; we recom­
mend replacing the batteries

Hide quick links: