► NOTICE! Only use the vacuum pump when it is installed on the adapter plate.
Use the selector switch to choose the desired mode. The vacuum pump reports certain oper­
ating states with signals.
Light signal
LEDs 1 and 2 flash white
LEDs 1 and 2 illuminated red
LEDs 1 and 2 flash blue
LEDs 1 and 2 flash green, every 10 s
LEDs 1 and 2 flash alternating blue
and red, every 10 s
LEDs 1 and 2 flash red

Leak warning information

The leak warning indicates that the desired vacuum cannot be achieved or maintained. The vacu­
um pump continues running. The leak warning is output under the following conditions:
The vacuum pump operates continuously for more than 3 minutes
Less than 2 minutes elapse between several pump intervals
The leak warning deactivates itself automatically when there is no problem for 10 minutes. It can
also be deactivated by briefly activating the vacuum pump standby mode.

6.3 Flushing the Vacuum Pump

A reduced vacuum indicates contamination. In case of contamination, the system has to be
flushed to clean it. This flushes the vacuum pump and the filter in the adapter plate. The reverse
mode of the vacuum pump is used for flushing. The flushing process can also help if the system is
plugged.
CAUTION
Improper cleaning
Skin irritation, formation of eczema or infections due to contamination with germs
Clean the product regularly.
Required materials:
>
Vessel with 30 ml to 60 ml distilled water, one collecting vessel
Preparations required:
Vacuum pump installed on adapter plate and in standby mode, prosthesis taken off the leg
1) Dip the outlet hose into the vessel with distilled water.
2) Hold the socket in such a way that the water runs into the collecting vessel during flushing.
3) Set the selector switch to reverse mode.
4) Allow the vacuum pump to run until the water has been pumped through the entire system.
5) Allow the pump to dry run for 15 s in order to pump out any remaining water.
6) Set the selector switch to standby mode.
7) Clean the interior surface of the socket.

7 Maintenance

► A visual inspection and functional test of the prosthetic components should be performed
after the first 30 days of use.
Standby mode
Signal tone, like siren (1x)
Selected mode
Audible signal
Signal tone rising (1x)
Signal tone falling (1x)
Click sound
Status
Restart of the device
Device switching to standby mode
Selector switch operated, mode
change
Pump running (forwards or in
reverse)
Device is powered up and working
without leaks
Leak warning (see information)
System error