Monitor Spectrum—Average Count
[:SENSe]:MONitor:AVERage:COUNt <integer>
[:SENSe]:MONitor:AVERage:COUNt?
Set the number of data acquisitions that will be averaged. After the
specified number of average counts, the averaging mode (terminal
control) setting determines the averaging action.
Factory Preset
and *RST:
10
Range:
1 to 1,000
Remarks:
You must be in the Modulation Analysis mode to use
this command. Use INSTrument:SELect to set the
mode.
Front Panel
Access:
Meas Setup
Monitor Spectrum—Averaging State
[:SENSe]:MONitor:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:MONitor:AVERage[:STATe]?
Turn averaging on or off.
Factory Preset
and *RST:
Off
Front Panel
Access:
Meas Setup
Monitor Spectrum—Averaging Termination Control
[:SENSe]:MONitor:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:MONitor:AVERage:TCONtrol?
Select the type of termination control used for the averaging function.
This determines the averaging action after the specified number of data
acquisitions (average count) is reached.
Exponential - After the average count is reached, each successive
data acquisition is exponentially weighted and combined with the
existing average.
Repeat - After reaching the average count, the averaging is reset and
a new average is started.
Chapter 7
Programming Language Reference
SENSe Subsystem
131