Programming Language Reference
SENSe Subsystem
Monitor Spectrum—Sweep Time Automation
[:SENSe]:MONitor:CHANnel:SWEep:TIME:AUTO ON|OFF|1|0
[:SENSe]:MONitor:CHANnel:SWEep:TIME:AUTO?
Automatically selects the fastest sweep time compatible with
measurement accuracy, or allows you to set it manually.
Factory Preset
and *RST:
On
Remarks:
You must be in the Modulation Analysis mode to use
this command. Use INSTrument:SELect to set the
mode.
Front Panel
Access:
Meas Setup, Chan Setup
Monitor Spectrum—Trigger Synchronization Control
[:SENSe]:MONitor:TRIGger:SOURce: IMMediate|EXTernal|RFBurst
[:SENSe]:MONitor:TRIGger:SOURce?
Provides the menu keys that allow you to select the source for the
demodulation synchronization trigger.
• Immediate - the next measurement is taken immediately, capturing
the signal asynchronously.
• External - sets the trigger directly to an external signal connected to
the rear-panel TRIGGER IN connector. No measurement will be
made unless an appropriate trigger signal is connected to the
Trigger In connector on the rear panel.
• RF Burst - sets the measurement trigger to the leading edge of a RF
Burst signal. No measurement will be made unless a bursted signal
is being measured.
Factory Preset
and *RST:
Immediate
Remarks:
You must be in the Modulation Analysis mode to use
this command. Use INSTrument:SELect to set the
mode.
Front Panel
Access:
Meas Setup, Trig Source
134