Preparing To Make Measurements; Initial Settings - Agilent Technologies ESA-E Series User And Programming Manual

Spectrum analyzers modulation analysis measurement personality
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Setting Up the Modulation Analysis Mode

Preparing to Make Measurements

Preparing to Make Measurements
At initial power up, the analyzer will be in spectrum analyzer (SA) mode and the
FREQUENCY Channel
measurement personality, press the
key.

Initial Settings

If you have already been into Modulation Analysis mode since the instrument was
powered up, then the screen displayed and the settings in force will be exactly as
they were when you last switched out of Modulation Analysis mode.
When you put the instrument into Modulation Analysis mode for the first time
after a power up, the Monitor Spectrum screen will be displayed with a center
frequency setting of 1.0 GHz and a span of 3 MHz. This Monitor Spectrum
function allows you to check that there is a signal to be measured.
The center frequency can be changed by pressing
required frequency using the numeric keys. The span can be altered by pressing
Span
and entering the required span using the numeric keys.
Before making a measurement, make sure the mode setup and radio standard
parameters are set to the desired settings. Refer to
Chapter 6 , "Front-Panel Key Reference,"
in changing parameter settings.
If you want to set the instrument (including the Error Vector Magnitude (EVM)
mode settings) to a known, factory default state, press
mode setup and all of the measurement setup parameters to the factory default
parameters. Note that
Preset
mode. You must re-access the EVM mode after the preset operation is completed.
You can set the instrument to use User preset or Factory preset under the
front panel key. If you set the preset to User, the instrument displays a Preset
menu when you press
Preset
defaults or the
Factory
using user defaults, refer to the ESA Spectrum Analyzers User's Guide.
If you want to set only the EVM mode to a known, factory default state, press
and
Mode Setup
Restore Mode Defaults
parameters to the factory defaults without affecting the SA mode, and the
instrument will not exit the EVM mode.
To preset only the settings that are specific to the selected measurement, press
Restore Meas Defaults
parameters, for the currently selected measurement only, to the factory defaults.
18
menu displayed. To access the Modulation Analysis
key and select the
Mode
for additional information to guide you
will switch modes, returning the ESA to the SA
. The Preset menu allows you to select the
defaults. For more information on setting, saving, and
. This will reset only the EVM
under
. This will set the measurement setup
Meas Setup
Modulation Analysis
Frequency
and then entering the
Chapter 5 , "Menu Maps,"
. This will preset the
Preset
User
Chapter 3
and
System