Factory Preset: ON
*RST:
ON (OFF recommended for remote operation)
Front Panel
Access:
Single
Meas Control, Measure Cont Single

Take New Data Acquisitions

:INITiate[:IMMediate]
This command initiates a sweep if not in a measurement. If in a
measurement and the measurement is in the idle state (i.e. not
running), it triggers the instrument if trigger conditions are met. If the
measurement is currently running, the command is ignored. A
"measurement" refers to any function under the MEASURE key.
Remarks:
See also the *TRG command and the TRIGger
subsystem.
Use the [:SENSe]:<meas>:TRIGger:SOURce command
to select the desired trigger. The instrument must be in
the single measurement mode. If
:INITiate:CONTinuous is ON then the command is
ignored.
Use :FETCh? to transfer a measurement result from
memory to the output buffer. Refer to individual
commands in the FETCh subsystem for more
information.
Front Panel
Access:
Sweep
Single
Meas Control
Chapter 7
Programming Language Reference
(SA mode)
INITiate Subsystem
109