Synchronization Commands - Agilent Technologies ESA-E Series User And Programming Manual

Spectrum analyzers modulation analysis measurement personality
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

PDC - Personal Digital Cellular
TETRA - Trans-European Trunked Radio
Factory Preset
and *RST:
CDMA
Front Panel
Access:
Mode Setup, Radio Std
Radio Standard Symbol Rate
[:SENSe]:RADio:STANdard:SRATe <symbol rate>
[:SENSe]:RADio:STANdard:SRATe?
Select the symbol rate of the radio signal being measured. EDGE
signals are limited to plus or minus 10%, but other signal standards
allow values up to 8 Msps. Calibration, however, is always limited to
the default value for each radio standard plus or minus 10%.
Factory Preset
and *RST:
1.22880 Msps
Maximum Value: 8 Msps
Minimum Value: 1 ksps
Front Panel
Access:
Det/Demod, Symbol Rate
or
Mode Setup, Demod...

Synchronization Commands

Burst Search Threshold
[:SENSe]:SYNC:BURSt:STHReshold <rel_power>
[:SENSe]:SYNC:BURSt:STHReshold?
Set the power threshold, relative to the peak power, that is used to
determine the burst rising edge and falling edge.
Factory Preset
−20 dB
and *RST:
−200 dB to −0.01 dB
Range:
Default Unit:
dB
Front Panel
Chapter 7
Programming Language Reference
SENSe Subsystem
139