Agilent Technologies ESA-E Series User And Programming Manual page 139

Spectrum analyzers modulation analysis measurement personality
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Error Vector Magnitude—Droop Compensation
[:SENSe]:EVM:DROop:COMPensation[:STATe] OFF|ON|0|1
[:SENSe]:EVM:DROop:COMPensation?
Sets Droop Compensation on or off.
Factory Preset
and *RST:
On
Remarks:
This key is only used with EDGE signals only. The
softkey is disabled for all other standards.
Front Panel
Access:
Meas Setup, More
Error Vector Magnitude - Symbol Dots (Quad View)
[:SENSe]:EVM:GSDOTs[:STATe] ON|OFF|1|0
[:SENSe]:EVM:GSDOTS[:STATe]?
Allows you to highlight the positions of the symbol decision points on
the graphical views by turning the display of the symbols on or off.
Factory Preset
and *RST:
On
Front Panel
Access:
Display
Error Vector Magnitude - I/Q Points Offset
[:SENSe]:EVM:IQOOffset <integer>
[:SENSe]:EVM:IQOOffset?
Specifies an offset from the first IQ point in the constellation diagram.
Factory Preset
and *RST:
0
Maximum Value: (Meas Interval × Points/Symbol) - I/Q Points
Minimum Value: 0
Front Panel
Access:
Display
Chapter 7
Programming Language Reference
,
Symbol Dots Quad View
SENSe Subsystem
125

Hide quick links: