Agilent Technologies ESA-E Series User And Programming Manual page 57

Spectrum analyzers modulation analysis measurement personality
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

2.
Figure 4-22
received symbol rate of 3.84150 Msps (this symbol rate can be seen
above the vector diagram in the figure.) The transmitted symbol rate
is 3.840 Msps.
Figure 4-22
Polar Vector Showing Symbol Rate Error
3.
Figure 4-23
1 that with a symbol rate error, the EVM versus time graph shows a
"V" shape.
Figure 4-23
I/Q Error Diagram Showing Symbol Rate Error
Chapter 4
Making Modulation Analysis Measurements
shows the polar vector diagram of this signal, with a
shows the I/Q error views of this signal. Observe at item
Interpreting Measurement Results
43

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4402bEsa-e e4404bEsa-e e4405bEsa-e e4407b

Table of Contents