Agilent Technologies ESA-E Series User And Programming Manual page 138

Spectrum analyzers modulation analysis measurement personality
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Programming Language Reference
SENSe Subsystem
Error Vector Magnitude—Averaging State
[:SENSe]:EVM:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:EVM:AVERage[:STATe]?
Turn average on or off.
Factory Preset
and *RST:
Off in front panel mode
On if you perform CONFigure:EVM
Front Panel
Access:
Meas Setup
Error Vector Magnitude—Averaging Termination Control
[:SENSe]:EVM:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:EVM:AVERage:TCONtrol?
Select the type of termination control used to averaging. This
determines the averaging action after the specified number of data
acquisitions (average count) is reached.
EXPonential – Each successive data acquisition after the average
count is reached, is exponentially weighted and combined with the
existing average.
REPeat – After reaching the average count, the averaging is reset
and a new average is started.
Factory Preset
and *RST:
Repeat
Front Panel
Access:
Meas Setup
Error Vector Magnitude—Burst Synchronization
[:SENSe]:EVM:BSYNc:SOURce RFAMplitude|NONE
[:SENSe]:EVM:BSYNC:SOURce?
Factory Preset
and *RST:
None
Front Panel
Access:
Meas Setup