7
Remote Programming
Operation Status Subsystem
110
The Operation Status register contains conditions which are part of the
instrument's normal operation.
Table 8
Bit Number
0
Not used
1
Not used
2
Not used
3
Not used
4
Arm Status
5
Trigger Status
6
Streaming Active
7
Memory Capture in Progress
8
Not used
9
Not used
10
ADC Calibration Run Status
11
Not used
12
Not used
13
Not used
14
Not used
15
Not used
Operation status register
Decimal Value
1
2
4
8
16
32
64
128
256
512
1024
2048
4096
8192
16384
32768
Keysight M8131A 16/32 GSa/s Digitizer User's Guide
Definition
Returns "0"
Returns "0"
Returns "0"
Returns "0"
Trigger is armed.
Trigger received.
Data streaming has been started and not
stopped yet.
Memory capture has been started and not
completed yet.
Returns "0"
Returns "0"
ADC calibration is running.
Returns "0"
Returns "0"
Returns "0"
Returns "0"
Returns "0"