Calibrate:adc:information; Calibrate:adc:recommended; Calibrate:adc:date[:Recommended] - Keysight Technologies M8131A User Manual

16/32 gsa/s
Hide thumbs Also See for M8131A:
Table of Contents

Advertisement

:CALibrate:ADC:INFormation?

Returned Format
Example

:CALibrate:ADC:RECommended?

Return Format

:CALibrate:ADC:DATE[:RECommended]?

Return Format
Keysight M8131A 16/32 GSa/s Digitizer User's Guide
Query
:CALibrate:ADC:INFormation?
Description
The query returns information about the last calibration of the ADC
module.
[:CALibrate:ADC:INFormation] <information_string><NL>
<information_string> Contains information about type ("User" or
"Factory") and date of last calibration.
Query
:CALibrate:ADC:INFormation?
"M8131A Setup:
Mode: 16GHz on 4 channels
Calibration Loaded:
1(User): 2018-03-09 17:33:44"
Query
:CALibrate:ADC:RECommended?
Parameters
None
Description
This query returns a Boolean value, indicating if an ADC calibration is
recommended.
[:CALibrate:ADC:RECommended] <recommended><NL>
Example
:CAL:ADC:REC?
Query
:CALibrate:ADC:DATE[:RECommended]?
Parameters
None
Description
This query returns the date when the next ADC calibration is
recommended.
[:CALibrate:ADC:DATE:RECommended] <date><NL>
Example
:CAL:ADC:DATE?
0
"2020-07-01"
Remote Programming
123