  • Page 1 Keysight M8131A 16/32 GSa/s Digitizer User's Guide...
  • Page 2 APPLICABLE LAW, KEYSIGHT DISCLAIMS under its standard commercial license, conditions are fully understood and ALL WARRANTIES, EITHER EXPRESS OR which is embodied in its End User License met. IMPLIED WITH REGARD TO THIS MANUAL Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 3: Safety Summary

    When the instrument is rack mounted the rack must be provided with an easily accessible mains switch. Line Power Requirements The Keysight M8131A operates when installed in an Keysight AXIe mainframe. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 4 (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 5: Instrument Markings

    The RCM mark is a registered trademark of the Australian Communications and Media Authority. The CSA mark is a registered trademark of the CSA International. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 6: Compliance And Environmental Information

    “Monitoring and Control instrumentation” product. Do not dispose in domestic household waste. To return unwanted products, contact your local Keysight office, or see http://about.keysight.com/en/companyinfo/environment/takeback.shtml for more information. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 7: Table Of Contents

    Front Panel Related Documents Additional Documents 2 M8131A Software Installation Pre-Requisites Installation Process Post Installation Steps How to use M8131A Instrument M8131A Maintenance ESD Protection Power and Ventilation Requirements Thermal Protection Battery Operating Environment Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 8 Acquisition Tab Trigger Tab Clock Tab Configurations Tab IOs Tab System Monitor Tab Working with M8131A Front Panel Numeric Control Usage Window Controls Usage Calibration Manager Exporting Waveforms Integrating 89600 VSA with M8131A Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 9 ADC Calibration Status Subsystem Filter Status Subsystem Sample Clock Status Subsystem ADC Calibration Run Status Subsystem Arm Status Subsystem Trigger Status Subsystem Memory Capture Status Subsystem Data Streaming Status Subsystem Acquire Commands :ACQuire: SRATe[?] :ACQuire:POINts[?] Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 10 :CALibrate:ADC:INFormation? :CALibrate:ADC:RECommended? :CALibrate:ADC:DATE[:RECommended]? :CALibrate:ADC:TEMPerature:DELTa[:CURRent]? :CALibrate:ADC:TEMPerature:DELTa:MAXimum? :CALibrate:ADC:TEMPerature:DELTa:MINimum? :CALibrate:ADC:SAVE :CALibrate:ADC:LOAD Latency Calibration Commands :CALibrate:SYNC:INPut:LEVel :CALibrate:LATency:LCMPeriod :CALibrate:LATency:MODE :CALibrate:LATency[:STEP] Channel Commands :CHANnel<N>:DISPlay :CHANnel<N>:OFFSet :CHANnel<N>:RANGe :CHANnel<N>:DPRocessing:DDC:CFRequency :CHANnel<N>:DPRocessing: CHARacteris[:VALue] :CHANnel<N>:DPRocessing:FILTer[:VALue] :CHANnel<N>:DPRocessing:FILTer:DEFault :CHANnel<N>:DPRocessing:FILTer:SCALe :CHANnel<N>:DPRocessing:FILTer:STATe Data Processing Commands :DPRocessing[:MODE] :DPRocessing:DDC:DECimation :DPRocessing:DIRect:DECimation Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 11 Contents Common Commands *IDN? *CLS *ESE ESR? *OPC *OPC? *OPT? *RST *SRE[?] *STB? *TST? *LRN? *WAI? Mass Memory Commands :MMEMory:CATalog? :MMEMory:CDIRectory :MMEMory:COPY :MMEMory:DELete :MMEMory:DATA :MMEMory:DATA:APPend :MMEMory:DATA? :MMEMory:MDIRectory :MMEMory:MOVE :MMEMory:RDIRectory :MMEMory:LOAD:CSTate :MMEMory:STORe:CSTate :MMEMory:STORe:WAVeform Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 12 :DIGitize :DIGitize:NDISplay :RUN :STOP :STReam :SSTReam System Commands :SYSTem:ERRor[:NEXT]? :SYSTem:HELP:HEADers? :SYSTem:LICense:EXTended:LIST? :SYSTem:SET[?] :SYSTem:VERSion? :SYSTem:COMMunicate:*? :SYSTem:COMMunicate:INSTr[:NUMBer]? :SYSTem:COMMunicate:HISLip[:NUMBer]? :SYSTem:COMMunicate:SOCKet[:PORT]? :SYSTem:COMMunicate:TELNet[:PORT]? :SYSTem:COMMunicate:TCPip:CONTrol? :SYSTem:ERRor:COUNT? Time Base Commands :TIMebase:REFClock Waveform Commands :WAVeform:BYTeorder :WAVeform:DATA? :WAVeform:DATA:BLOCk? :WAVeform:XINCrement? :WAVeform:XORigin? :WAVeform:YINCrement? :WAVeform:YORigin? Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 13 Contents Instrument Commands :INSTrument:SLOT[:NUMBer]? :INSTrument:IDENtify :INSTrument:IDENtify:STOP :INSTrument:HWRevision? :INSTrument:FPGA:MODE :INSTrument:FPGA:UPDate :INSTrument:FPGA:RESet :INSTrument:FPGA:DIRect:DWORd :INSTrument:AINPuts? :INSTrument:AMEMory? :INSTrument:MONitor:CURRent[:TOTal]? :INSTrument:MONitor:TEMPerature? Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 14 Contents Optical Data Interface Commands :ODI:ACHannels? :ODI:PORT:COUNt? :ODI:PORT<N>:CAPability:DIRection? :ODI:PORT[N]:CAPability:FCONtrols? :ODI:PORT[N]:CAPability:LANes? :ODI:PORT[N]:CAPability:NAME? :ODI:PORT[N]:CAPability:RATes? :ODI:PORT[N]:CAPability:RBMax? :ODI:PORT[N]:CAPability:TBMax? :ODI:PORT[N]:CAPability:TRMatch? :ODI:PORT[N]:NAME? :ODI:PORT<N>:ACTivate :ODI:PORT<N>:DEACtivate :ODI:PORT<N>:CSTatus? :ODI:PORT[N]:PSTatistics:BBURsts? :ODI:PORT[N]:PSTatistics:RBYTes? :ODI:PORT[N]:PSTatistics:TBYTes? :ODI:PORT[N]:PSTatistics:THOFfs? :ODI:PRODucer:COUNt? :ODI:PRODucer[N]:CAPability:CLIDs? :ODI:PRODucer<N>:CAPability:CCIDs? :ODI:PRODucer<N>:CAPability:NAME? :ODI:PRODucer<N>:CAPability:PFORmats? :ODI:PRODucer<N>:NAME? :ODI:PRODucer<N>:IFSupported :ODI:PRODucer<N>:ACTivate :ODI:PRODucer<N>:DEACtivate Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 15 Contents Trigger Commands : TRIGger:SWEep : TRIGger:EDGE:SOURce : TRIGger:EDGE:SLOPe :TRIGger:LEVel:EXTernal :TRIGger:LEVel[:CHANnel] :TRIGger:HOLDoff : TRIGger:HOLDoff:TYPE Test Commands :TEST:PON? :TEST:TST? 5 Characteristics Performance Specification Operating Environment General Index Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 17: Introduction

    This chapter provides an overview of M8131A digitizer module. M8131A Overview The Keysight M8131A is a 16/32 GSa/s digitizer with high bandwidth and channel density. It offers up to 1.6 GSa waveform memory. The M8131A is ideally suited to address following key applications: •...
  • Page 18: Feature Options

    On board memory 1.6 GSa per module • Form factor: 2-slot AXIe module controlled via external PC or embedded AXIe system controller M9536A Feature Options The following feature options are available for M8131A: Table 3 M8131A feature options Option Functionality...
  • Page 19 Introduction Figure 1 on page 19 illustrates the front panel of the M8131A instrument. Figure 1 M8131A front panel The M8131A front panel include the following input/output ports: Inputs/Outputs • Input - The four analog input ports (In 1/2/3/4) can be used to receive the analog data.
  • Page 20 It can be configured in two independent groups of five bits as an input or output bus. If option FP1 is installed, the Control In/Out is routed to the ‘open FPGA’. The Control In/Out functionality is for future use and is currently not supported. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 21: Related Documents

    In case the output relay has shut-off because of an external overload condition, this LED flashes. Related Documents To access documentation related to the Keysight M8131A digitizer, use one of the following methods: • CD - Browse the product CD for M8131A documentation.
  • Page 22: Additional Documents

    Additional documentation can be found at: • http://www.keysight.com/find/M9502A for 2-slot chassis related documentation. • http://www.keysight.com/find/M9505A for 5-slot chassis related documentation. • http://www.keysight.com/find/M9514A for 14-slot chassis related documentation. • http://www.keysight.com/find/M9537A for embedded AXIe controller related documentation Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 23: M8131A Software Installation

    Keysight M8131A 16/32 GSa/s Digitizer User’s Guide M8131A Software Installation / 24 Pre-Requisites / 25 Installation Process / 35 Post Installation Steps / 37 How to use M8131A Instrument / 38 M8131A Maintenance This chapter explains the steps required to install M8131A software package.
  • Page 24: Pre-Requisites

    M8131A Software Installation Pre-Requisites The following are the pre-requisites for installing Keysight M8131A software: The supported operating systems are: • Windows 10 (32 bit or 64 bit) • Windows 8.1 (32 bit or 64 bit) • Windows 7 (32 bit or 64 bit) Ensure that you have Keysight IO Libraries Suite 2017 Update 1 or higher installed on your system.
  • Page 25: Installation Process

    M8131A Software Installation Installation Process Follow the given steps to install Keysight M8131A software on your system: Double-click the executable (M8131_Setup.exe). This executable file is available either on CD or Web. The installer will first check and list some pre-requisites. Click Install to NOTE install them.
  • Page 26 M8131A Software Installation Follow the on-screen instructions to begin the installation process. Click Next. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 27 We recommend you to read the document to check if your hardware configuration is supported. Click Next to proceed to the license agreements. Accept the terms of Keysight software end-user license agreement and click Next. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 28 M8131A Software Installation Select Yes if you want to read the post-installation instructions when finished. Click Next to select setup type. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 29 M8131A Software Installation Select a setup type either Complete or Custom. Click Next. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 30 M8131A Software Installation If you select Custom and click Next, you can specify which optional features will be installed: Click Next to begin installation. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 31 M8131A Software Installation 10 The Setup Wizard will now install M8131 beginning with the kernel driver installation. If you have clicked “Always trust Keysight” before you won’t get this dialog anymore. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 32 M8131A Software Installation 11 The Keysight M8131A will configure the new software installation. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 33 M8131A Software Installation The following screen will appear once the Keysight M8131A software is successfully installed on your system. Click Finish to restart your system. Do not connect the AXIe chassis to your system using the PCIe or USB cable during this reboot.
  • Page 34 M8131A Software Installation 12 This completes the Keysight M8131A software installation. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 35: Post Installation Steps

    Post Installation Steps Follow the post installation steps as shown below: If M8131A is already powered up and connected to PC using the PCIe, NOTE just reboot the PC and start with step 5. No such reboot step is required in case of USB connection.
  • Page 36 Choose what the power buttons do 5 Under the pane, Change settings that are currently unavailable ensure that is not selected. Turn on fast startup The steps above may vary depending on your Windows version. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 37: How To Use M8131A Instrument

    How to use M8131A Instrument In order to use the instrument: If you use a PCIe link to control the M8131A, the AXIe chassis must be switched on before you start the PC. If you use an USB link to control the M8131A, it’s not mandatory that the AXIe chassis is powered and...
  • Page 38: M8131A Maintenance

    CAUTION There are also several exposed components on the PCAs, on both sides of M8131A, which can be touched accidentally while handling the unit and can risk damage to the instrument, due to ESD. When you connect a device or cable that is not fully discharged to these connectors, you risk damage to the instrument and expensive instrument repairs.
  • Page 39 1 MW of isolation from ground. These techniques for a static-safe work station should not be used when WARNING working on circuitry with a voltage potential greater than 500 volts. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 40: Power And Ventilation Requirements

    Keysight Service for repair. Battery The M8131A does not have a battery. Operating Environment For details on operative environment for M8131A, refer to the section Operating Environment on page 153. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 41: M8131A Soft Front Panel

    Configurations Tab / 69 IOs Tab / 71 System Monitor Tab / 72 Working with M8131A Front Panel / 74 Calibration Manager / 78 Exporting Waveforms / 79 Integrating 89600 VSA with M8131A This chapter describes the M8131A Soft Front Panel.
  • Page 42: Launching The M8131A Soft Front Panel

    This dialog shows the addresses of the discovered M8131A modules. Select a module from the list and press “Connect”. If no M8131A module is connected to your PC, you can select “Simulation Mode” to simulate an M8131A module. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 43 M8131A User Interface Figure 5 M8131A in simulation mode Next, a M8131A software startup screen will be displayed as shown in Figure 6 on page 44. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 44 M8131A User Interface Figure 6 M8131A startup screen Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 45: Command Line Arguments

    Additional information shown in the SFP window title. /OutputDir Set the output directory for the log file and temporary files. /r resourceName Visa PXI resource string of the module to connect to, e.g. PXI12::0::0::INSTR. “auto” selects the next free instrument. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 46: Communication

    The socket and telnet port are then calculated from the HiSLIP index: • telnet port = 60000 + 2 * <HiSLIP index> • socket port = 60000 + 2 * <HiSLIP index> + 1 Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 47 /Socket, /Telnet, /Inst, /HiSLIP or their respective default values instead. If both /NoAutoID and /AutoID are specified, /AutoID overrides /NoAutoID. The first port not assigned by IANA is 49152 (IANA, Internet Assigned NOTE Numbers Authority, http://www.iana.org Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 48: M8131A Soft Front Panel

    M8131A User Interface M8131A Soft Front Panel The M8131A Soft Front Panel and its GUI elements are illustrated in Figure 7 on page 48. Figure 7 M8131A user interface The M8131A Soft Front Panel includes the following GUI elements: Title Bar...
  • Page 49: Title Bar

    File > Export Waveform Exports the data of a channel to a text file with a header. See :MMEMory:STORe:WAVeform command for a description of the file format. File > Exit Exits the M8131A application. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 50 Not functional in the current software release. Trace Menu The Trace menu includes the following selections: Trace > Time Window Opens a new instance of time window. Trace > Frequency Window Opens a new instance of frequency window. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 51 Tools > System Monitor Switch to the “System Monitor” tab on Parameters window if it is already System open. If not, it adds “System Monitor” tab first. For details, see Monitor Tab on page 71. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 52 Windows > Tile Horizontally Aligns the multiple opened windows in a horizontal sequence. The following figure shows the horizontal sequence of two different windows. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 53 Help > Online Support Opens the instrument’s product support web page. Help > About Displays product information including version number, build date, build info, installed licenses, available options and web links for M8131A information and support. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 54: Tool Bar

    Auto Scale adjusts All open windows’ scaling so that the whole traces are visible. An auto-scale for one window can be done via Auto Scale inside graph panel’s context menu (i.e. right-click and use context menu). Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 55: Lower Pane

    Opens the Status Window. This feature is currently not implemented. Show Error List Window Opens the window that shows the list of errors and warnings. For details, see Errors List Window on page 56. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 56: Errors List Window

    Click this toggle button to automatically open the errors list window whenever an error occurs (default) or not. Copy Click this button to copy the selected message(s). Select All Click this button to select all messages inside the list. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 57: Status Bar

    Parameters Window The Parameters Window provides various tabs which are used to configure the parameters of the M8131A instrument. These tabs include: Acquisition, Trigger, Clock, Configurations, IOs and System Monitor. Figure 9 on page 58 illustrates an example of the Parameter Window.
  • Page 58 Parameters window You can use the numeric controls to adjust the values and units. For NOTE details on how to use numeric control, refer to the section Numeric on page 72. Control Usage Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 59: Trace Window

    M8131A User Interface Trace Window The trace window contains time and frequency graph panels. A graph panel is an area on which the plotted graphs for a signal is shown. Figure 10 Frequency graph Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 60 Figure 11 Time graph The context-menu of a graph panel provides features (i.e., effective on the current graph panel and not on the other open panels): Figure 12 Context menu of a graph Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 61 • Show Grid: Use this check-box to show/hide grids from the graph. • Clear Marker: It clears the selected marker from the graph panel. It is only enabled when a marker is selected. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 62: Acquisition Tab

    Sample Rate - Based on available options, a sample rate of 16 GSa/s or 32 GSa/s can be selected. • Mode - The following two modes are available: • Direct Mode - The direct mode is available for both 16 GSa/s or 32 GSa/s sample rates. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 63 Display On: Turns on or off the waveform memories for corresponding channel number and display its content in trace panel(s). The “Run” capture button is functional only when at least one channel is selected. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 64 DDC IF values; 0 MHz, 250 MHz, 500 MHz, 750 MHz, 1 GHz & 1.25 GHz. • Frontend Settings - The forntend settings are available, depending upon the installed options (see Instrument Options on page 18). These settings allow you to set the signal offset and range. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 65: Trigger Tab

    M8131A User Interface Trigger Tab The Trigger tab provides the trigger settings to M8131A module. Figure 14 Trigger tab Settings • Source - Select the trigger source for sweep mode triggering. The following options are available: • Trigger Input - Signal acquisition starts when a positive/negative slope is received on the trigger input port of M8131A.
  • Page 66 Holdoff Type - Select an option from the Holdoff Type drop-down list. The following options are available: • Normal - Once the hold off time is elapsed, the M8131A waits for a trigger and then it triggers only after the trigger is received. •...
  • Page 67: Clock Tab

    Input Clock Settings • Ref. Clock Source - A clock reference input is provided on the front panel of the M8131A module. It is used as the clock reference for all four analog channels of that M8131A. The options are: •...
  • Page 68: Configurations Tab

    Select a FGPA setting files in the module (*.bit) and then click Open. Click on the Configure button to load the selected file. Click on the Default button to set the FPGA setting back to the default state. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 69: Ios Tab

    Enable Port - The respective ODI ports can be activated/deactivated by selecting the corresponding check-boxes. If these check-boxes are selected without making any physical connections, an error for loss of signal will be shown. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 70 None - No flow control. • In Band - In-band flow control. • Statistics - Display the ODI statistics in terms of Status, Bytes Sent, Bytes Received, Bad Bursts Received, and Tx Flow Holdoffs. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 71: System Monitor Tab

    Figure 18 System Monitor tab • Module Total Current: The measured current can be updated by pressing the Refresh button. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 72: Working With M8131A Front Panel

    For example, in the Frequency control you can use "H", "K", "M", or "G" to specify Hz, KHz, MHz or GHz , respectively. (The control is not case sensitive.) Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 73: Window Controls Usage

    For example, you can select and generate various layouts from the single window. You can move a window anywhere on the screen or to a different monitor. You can also close the floating windows. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 74: Calibration Manager

    • Differential inputs: One calibration step, start it and wait until it is finished. Figure 21 Calibrated zero volts on all channels with 50-ohm resistor caps installed Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 75 M8131A User Interface The calibration dialog is used to calibrate M8131A module. You can also save and load the calibration for future use. Figure 22 Calibration dialog Starting Calibration Process To start the calibration: Go to Utilities > Calibration. The Calibration Manager dialog box appears.
  • Page 76 To save the calibration; Click on the Save button. A standard Windows Save As dialog box will appear. Figure 23 Saving calibration Specify the name (.TXT) and location Click Save. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 77 M8131A User Interface To load the calibration; Click on the Load button. A standard Windows Open dialog box will appear. Figure 24 Loading calibration Locate the calibration file (.TXT). Click Open. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 78: Exporting Waveforms

    Go to File > Export Waveform and then select a channel. A standard Windows Save Waveform to File dialog box will appear. Figure 25 Exporting waveforms Specify the name (.TXT) and location Click Save. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 79: Integrating 89600 Vsa With M8131A

    M8131A User Interface Integrating 89600 VSA with M8131A Follow the given steps to configure M8131A with 89600 VSA: Get the HiSLIP information a Start the M8131A Soft Front Panel. b On the menu bar, click Help > About. c Copy the HiSLIP information from the dialog.
  • Page 80 Click the Enter Address tab. Enter the LAN address or hostname, select the protocol used to communicate with the instrument. g Click OK. Integrate 89600 VSA with M8131A a Start the 89600 VSA application. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 81 New Hardware Configuration dialog box will appear: e Select M8131 and click [>] Rename the configuration as per your choice and click OK. g Select the new Analyzer Configuration (M8131A(1)) as shown in the following figure: Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 82 M8131A User Interface h The 89600 VSA will now acquire the M8131A. Close the Hardware dialog and start the measurement. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 83: Remote Programming

    Keysight M8131A 16/32 GSa/s Digitizer User’s Guide Remote Programming / 84 Remote Programming Overview / 86 Status Commands / 98 Acquire Commands / 100 ADC Calibration Commands / 104 Latency Calibration Commands / 106 Channel Commands / 111 Data Processing Commands...
  • Page 84: Remote Programming Overview

    Remote Programming Remote Programming Overview This chapter introduces the basics for remote programming of an M8131A instrument using SCPI commands. Instructions Instructions, both commands and queries, normally appear as strings embedded in a statement of your host language, such as Visual Basic for Applications (VBA), Visual Basic .NET, C#, C, etc.
  • Page 85: Braces

    When there is more than one data parameter, they are separated by commas (,). You can add spaces around the commas to improve readability. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 86: Status Commands

    Status Commands This section describes the structure of the SCPI status system used by the M8131A. The status system records various conditions and states of the instrument in several register groups as shown on the following pages. Each of the register groups is made up of several low level registers called Condition registers, Event registers, and Enable registers which control the action of specific bits within the register group.
  • Page 87 ADC Calibration Run Status Standard Event Status Register Status Byte Operation Complete Query Error Device Dependent Error Execution Error Command Error Power On Summary of IEEE 488.2 Status Structure Registers Figure 26 Status register structure Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 88: Status:preset

    One or more bits are set in the Standard Event Register Master Summary One or more bits are set in the Status Byte Register Operational Data One or more bits set in the Operation Data Register (bits must be enabled) Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 89: Questionable Data Register Command Subsystem

    For more details, see Status Subsystem on page 92. Not used 2048 Returns “0” Not used 4096 Returns “0” Not used 8192 Returns “0” Not used 16384 Returns “0” Not used 32768 Returns “0” Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 90: Operation Status Subsystem

    ADC Calibration Run Status 1024 ADC calibration is running. Not used 2048 Returns “0” Not used 4096 Returns “0” Not used 8192 Returns “0” Not used 16384 Returns “0” Not used 32768 Returns “0” Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 91: Reference Clock Status Subsystem

    Frequency of external reference signal out-of-range. Connection Status Subsystem The Connection Status register contains the state of the USB connection to the M8131A module. The following SCPI commands and queries are supported: :STATus:QUEStionable:CONNection[:EVENt]? :STATus:QUEStionable:CONNection:CONDition? :STATus:QUEStionable:CONNection:ENABle[?] :STATus:QUEStionable:CONNection:NTRansition[?] :STATus:QUEStionable:CONNection:PTRansition[?] Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 92: Adc Calibration Status Subsystem

    ADC calibration was aborted by the user. Calibration failed ADC calibration failed. Upon software start, bit 2 is set and bits 3, 4 are cleared. When a NOTE calibration was run, the bits are updated accordingly. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 93: Filter Status Subsystem

    A sample with an ADC code outside the valid range of -511 to +511 was detected after filtering at the Channel 2 Filter output clipped filter output Channel 3 Filter output clipped Channel 4 Filter output clipped Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 94: Sample Clock Status Subsystem

    Decimal Value Definition Sample Clock invalid No sample clock signal is connected to Sample Clk In, or the connected sample clock is invalid and cannot be used, because its amplitude or frequency is out-of-range. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 95: Adc Calibration Run Status Subsystem

    Bit 1 is set while the calibration for a single channel is running. A ‘0’ indicates that the calibration probe can be connected to the next channel and the :CALibrate:ADC:NEXT SCPI command can be issued again. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 96: Arm Status Subsystem

    Trigger Status Subsystem The following SCPI commands and queries are supported: :STATus:OPERation:TRIGgered[:EVENt]? :STATus:OPERation:TRIGgered:CONDition? :STATus:OPERation:TRIGgered:ENABle[?] :STATus:OPERation:TRIGgered:NTRansition[?] :STATus:OPERation:TRIGgered:PTRansition[?] Table 16 Trigger status register Bit Number Decimal Value Definition Trigger Status A trigger was detected. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 97: Memory Capture Status Subsystem

    The following SCPI commands and queries are supported: :STATus:OPERation:STReam[:EVENt]? :STATus:OPERation:STReam:CONDition? :STATus:OPERation:STReam:ENABle[?] :STATus:OPERation:STReam:NTRansition[?] :STATus:OPERation:STReam:PTRansition[?] Table 18 Data Streaming status register Bit Number Decimal Value Definition Data Streaming in Data streaming has been started and not stopped Progress yet. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 98: Acquire Commands

    (Direct/DDC) and the decimation factor. If the passed parameter does not fulfill the granularity the value is adjusted to the next higher supported value. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 99 268435456 Direct, Decimation by 4 268435456 134217728 Sample Rate 32 GSa/s Minimum Maximum Granularity Direct, No Decimation 838860800 Query :ACQuire:POINts? The query returns the value of the memory depth. Returned Format [:ACQuire:POINts] <points_value><NL> Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 100: Adc Calibration Commands

    This command applies only to the modules with single ended inputs, NOTE where a calibration requires an external calibration probe. :CALibrate:ADC:ABORt Command :CALibrate:ADC:ABORt Description The command aborts a running self-calibration of the ADC module. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 101: Calibrate:adc:information

    Return Format [:CALibrate:ADC:RECommended] <recommended><NL>  Example :CAL:ADC:REC? :CALibrate:ADC:DATE[:RECommended]? Query :CALibrate:ADC:DATE[:RECommended]? Parameters None Description This query returns the date when the next ADC calibration is recommended. Return Format [:CALibrate:ADC:DATE:RECommended] <date><NL>  Example :CAL:ADC:DATE? “2020-07-01” Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 102: Calibrate:adc:temperature:delta[:Current]

    :CALibrate:ADC:TEMPerature:DELTa:MINimum? Parameters None Description This query returns the minimum ADC temperature difference to the last ADC calibration before a new calibration is recommenced. Return Format [:CALibrate:ADC:TEMPerature:DELTa:MAXimum?] <min temperature difference><NL>  Example :CAL:ADC:TEMP:DELT? -1.00000000000000E+01 Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 103: Calibrate:adc:save

    <file_name> A quoted ASCII string with a maximum of 254 characters including the entire path name, if used. The filename assumes the present working directory, if a path does not precede the file name. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 104: Latency Calibration Commands

    • COMBined - Phase alignment to the master’s core clock and latency calibration are done in the same step. Query :CALibrate:LATency:MODE? Description This query returns the current latency calibration mode. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 105: Calibrate:latency[:Step]

    FIFOs are adjusted accordingly. • PREPare - Set the SYNC OUT to pulse mode (master, SEParate). When data streaming is started, a single pulse is sent at the SYNC OUT. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 106: Channel Commands

    The query returns the current offset value for the specified channel. Returned Format [:CHANnel<N>:OFFSet] <offset><NL> In modules with differential inputs, the query always returns NOTE 0.00000000000000E+00 and setting any value not equal to 0 results in an error. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 107: Channel:Range

    The valid range is from 0 to 1250 MHz in steps of 250 MHz. Query :CHANnel<N>:DPRocessing:DDC:CFRequency? The query returns the current center frequency of the DDC for the specified channel. Returned Format [:CHANnel<N>:DPRocessing:DDC:CFRequency] <frequency_value><NL> Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 108: Channel:Dprocessing: Characteris[:Value]

    The correction filter is applied directly after the conversion by the ADC and before further processing, for example by the DDC. <N> An integer, 1-4 in a single digitizer system. S16G Selects the coefficients for 16GSa/s S32G Selects the coefficients for 32GSa/s Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 109: Channel:Dprocessing:filter:default

    If the optional parameter of sample rate mode is given, it returns the default filter values for the specified sample rate otherwise for the current sample rate mode. Returned Format [:CHANnel<N>:DPRocessing:FILTer:DEFault] <filter_values> <NL> Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 110: Channel:Dprocessing:filter:scale

    The command sets the state of the correction filter - on or off. Query :CHANnel<N>:DPRocessing:FILTer:STATe? The query returns the current state of the correction filter for a channel. Returned Format [:CHANnel<N>:DPRocessing:FILTer:STATe] {1 | 0} <NL> Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 111: Data Processing Commands

    <N> An integer, 1-4 in a single digitizer system. POW5 2 POW6 2 POW7 2 POW8 2 Query :CHANnel<N>:DPRocessing:DDC:DECimation? The query returns the current decimation factor of the DDC for the specified channel. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 112: Dprocessing:direct:decimation

    All decimation factors are powers of 2. Parameters POW0 2 POW1 2 POW1 2 Example Command :DPR:DIR:DEC POW1 Query :DPR:DIR:DEC? -> POW1 The query returns the current decimation factor of the Direct mode for the specified channel. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 113: Common Commands

    A query of this register returns a decimal value which corresponds to the binary-weighted sum of all bits set in the register. *OPC Set the “Operation Complete” bit (bit 0) in the Standard Event register after the previous commands have been completed. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 114: Opc

    Poll but the “Master Summary” bit (bit 6) is not cleared by the *STB? command. *TST? Execute Self Tests. If self-tests pass, a 0 is returned. A number lager than 0 indicates the number of failed tests. To get actual messages, use :TEST:TST? Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 115: Lrn

    Use :SYST:SET to send the learn string. See :SYSTem:SET[?] on page 127. *WAI? Prevents the instrument from executing any further commands until the current command has finished executing. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 116: Mass Memory Commands

    <file_type> is always empty. <file_size> provides the size of the file in bytes. In case of directories, <file_entry> is surrounded by square brackets and both <file_type> and <file_size> are empty. Example Query :MMEM:CAT? Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 117: Mmemory:cdirectory

    The first pair of parameters specifies the source. The second pair specifies the destination. An error is generated if the source doesn't exist or the destination file already exists. Parameters <string>,<string> Example Command :MMEM:COPY "C:\data.txt", "C:\data_new.txt" Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 118: Mmemory:delete

    Description The command form is MMEMory:DATA:APPend <file_name>,<data>. It appends <data> to the existing file <file_name>. <data> is in 488.2 block format. <file_name> is string data. Parameters <file_name>, <data> Example Command :MMEM:DATA:APP "C:\data.txt", #14test Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 119: Mmemory:data

    The first pair of parameters specifies the source. The second pair specifies the destination. An error is generated if the source doesn't exist or the destination file already exists. Parameters <string>,<string> Example Command :MMEM:MOVE "C:\data_dir","C:\newdata_dir" Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 120: Mmemory:rdirectory

    Current state of instrument is loaded from a file. Parameters <file_name> Example Command :MMEM:LOAD:CST "C:\data.txt" :MMEMory:STORe:CSTate Command :MMEMory:STORe:CSTate <file_name> Description Current state of instrument is stored to a file. Parameters <file_name > Example Command :MMEM:STOR:CST "C:\data.txt" Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 121: Mmemory:store:waveform

    OFF | 0 Only data without header (default). Example: File header and beginning of data for a waveform captured in DIRect mode: Revision: 0 Type: Points: 1600 XInc: 6.25e-11 XOrg: XUnits: second YInc: 0.0005 YOrg: YUnits: Volt Data: 3.23486E-02 3.33252E-02 Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 122 Example: File header and beginning of data for a waveform captured in DDC mode: Revision: 0 Type: Points: 1600 XInc: 6.25e-11 XOrg: XUnits: second YInc: 0.0005 YOrg: YUnits: Volt Data: 1.23486E-02, 1.33252E-02 1.25348E-02, 1.39842E-02 Example Command :MMEM:STORe:WAVeform CHANnel1, "waveform.txt", TXT, Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 123: Root Level Commands

    Command execution can be stopped with the :STOP command. :STOP Command :STOP The command causes the digitizer to stop acquiring data into capture memory. It can be used to abort a :DIGitize, :DIGitize:NDISplay or :RUN command. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 124: Stream

    This command is only available with option STR. It stops ODI streaming on all ports. Example: Stop the data streaming on ODI port 1. // Stop ODI streaming. :SSTR // De-activate the data producer. :ODI:PROD1:DEAC Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 125 Remote Programming // De-activate the data consumer on the other side of the ODI link. // De-activate ODI port. :ODI:PORT1:DEAC // De-activate ODI port on the other side of the ODI link. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 126: System Commands

    The response is a <DEFINITE LENGTH ARBITRARY BLOCK RESPONSE DATA> element. The full path for every command and query is returned separated by linefeeds. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 127: System:license:extended:list

    In set form, the block data must be a complete instrument set-up read using the query form of the command. This command has the same functionality as the *LRN command. Parameters <binary block data> Example Command :SYST:SET <binary block data> Query :SYST:SET? Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 128: System:version

    Then this socket port can be queried using the HiSLIP connection. Example Query :SYST:COMM:*? :SYSTem:COMMunicate:INSTr[:NUMBer]? Command :SYSTem:COMMunicate:INSTr? Description The query returns the VXI-11 instrument number used by the Soft Front Panel. Example Query :SYST:COMM:INST? Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 129: System:communicate:hislip[:Number]

    Command :SYSTem:COMMunicate:TCPip:CONTrol? Description The query returns the port number of the control connection. You can use the control port to send control commands (for example “Device Clear”) to the instrument. Example Query :SYST:COMM:TCP:CONT? Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 130: System:error:count

    Remote Programming :SYSTem:ERRor:COUNT? Command :SYSTem:ERRor:COUNT? Parameters None Description This query returns the error count. Examples Query :SYST:ERR:COUNT? -> “5” Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 131: Time Base Commands

    Query :TIMebase:REFClock? The query returns the currently selected reference clock source. Returned Format [:TIMebase:REFClock] {INTernal | E010 | E100}<NL> Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 132: Waveform Commands

    16-bit integer representation are transferred. <N> An integer, 1-4. <start> An integer value which is the starting point in the source memory which is the first waveform point to transfer. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 133: Waveform:data:block

    [:WAVeform:DATA:BLOCk] {binary_block_data}<NL> :WAVeform:XINCrement? Command :WAVeform:XINCrement? CHANnel<N> Description This query returns the time difference between consecutive data points for the channel given by the first parameter. <N> An integer, 1-4. Example :WAVeform:XINCrement? CHAN1 -> 6.25000000000000E-11 Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 134: Waveform:xorigin

    This query returns the y-origin voltage value for the channel given by the first parameter. The voltage value returned is the voltage value represented by the waveform data digital code zero. <N> An integer, 1-4. Example WAVeform:YORigin? CHAN1 -> 0.00000000000000E+00 Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 135: Instrument Commands

    Remote Programming Instrument Commands The INSTrument subsystem contains queries to get information like occupied AXIe slot number and M8131A hardware revision. Additionally, if option -FP1 is installed, commands to update DPU FPGA images and to access DPU FPGA registers are available.
  • Page 136: Instrument:fpga:mode

    <file_name> File name of the FPGA update image :INSTrument:FPGA:RESet Command :INSTrument:FPGA:RESet {DPUA|DPUB} Description The command loads the default image into the DPU FPGA. DPUA Selects DPU A for update. DPUB Selects DPU B for update. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 137: Instrument:fpga:direct:dword

    The query returns the usable capture memories as a string. The memory names are separated by commas. The returned value depends on the installed options and the selected sample rate. Returned Format [:INSTrument:AMEMory] {active_channels_string}<NL> Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 138: Instrument:monitor:current[:Total]

    The query returns the temperature for the selected component. ADC Current ADC temperature LACAl Temperature of last ADC calibration FA Temperature of FPGA A FA Temperature of FPGA B Returned Format [:INSTrument:MONitor:TEMPerature] <value> Example :INST:MON:TEMP? ADC -> 7.09999990463257E+01 Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 139: Optical Data Interface Commands

    Optical Data Interface Commands The Optical Data Interface subsystem is used to setup the optical ports of the M8131A. These commands are only available with option -STR. An example sequence of commands to set up the ODI ports for data...
  • Page 140: Odi:port[N]:Capability:fcontrols

    The query returns the number of lanes supported by this port. Return Format [:ODI:PORT[N]:CAPability:LANes] <lanes><NL> Example :ODI:PORT1:CAP:LAN? -> 12 :ODI:PORT[N]:CAPability:NAME? Query :ODI:PORT[N]:CAPability:NAME? Parameters None Description The query returns the name of the port. Return Format [:ODI:PORT[N]:CAPability:NAME] <name><NL> Example :ODI:PORT:CAP:NAME? -> “ODI1” Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 141: Odi:port[N]:Capability:rates

    Example :ODI:PORT:CAP:RBM? -> 2048 :ODI:PORT[N]:CAPability:TBMax? Query :ODI:PORT[N]:CAPability:TBMax? Parameters None Description The query returns the supported transmitter maximum burst values. Return Format [:ODI:PORT[N]:CAPability:TBMax] <burst-max-list><NL> <burst-max-list> ::= [256 | 2048] [,…] Example ODI:PORT:CAP:TBMax? -> 256,2048 Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 142: Odi:port[N]:Capability:trmatch

    <N> An integer to select the ODI port, 1-4. <lane_rate> Lane rate R141 14.1 Gbit/s. <tx_burst_max> Maximum transmit burst size in bytes. Possible values are 256 and 2048 <directionality> Directionality PRODucer Transmit direction CONSumer Receive direction <tx_flow_control> Transmit flow control NONE Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 143: Odi:port:Deactivate

    The query returns the communication status of an optical port. <N> An integer to select the ODI port, 1-4. Returned Format [:ODI:PORT<N>:CSTatus] <status><NL> <status> 32-bit integer, meaning of the status bits is described in Table 19 on page -144. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 144: Odi:port[N]:Pstatistics:bbursts

    :ODI:PORT<N>:CSTatus? query. All status bits will be 0 on an inactive port. :ODI:PORT[N]:PSTatistics:BBURsts? Query :ODI:PORT[N]:PSTatistics:BBURsts? Parameters None Description This query returns the bad burst received by the ODI port. Example :ODI:PORT1:PST:BBUR? -> <integer> Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 145: Odi:port[N]:Pstatistics:rbytes

    Description This query returns the transmission holdoffs. Example :ODI:PORT1:PST:THOF? -> <integer> :ODI:PRODucer:COUNt? Command :ODI:PRODucer:COUNt? Parameters None Description This query returns the number of current producer count in the collection. Example :ODI:PORD:COUN? -> 4 Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 146: Odi:producer[N]:Capability:clids

    This query returns the name of the producer. Example :ODI:PROD1:CAP:NAME? -> "Ch1 DDC to ODI1" :ODI:PRODucer<N>:CAPability:PFORmats? Command :ODI:PRODucer<N>:CAPability:PFORmats? Parameters None Description This query returns the name of the packet formats available. Example :ODI:PROD1:CAP:PFOR? -> "NHEader" Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 147: Odi:producer:Name

    <packet_format> Specifies format of packet header. NHEader Raw binary samples with no header. <class_id> Specifies the binary data format of the packet payload. IQ16B1CH Complex (IQ), 16-bit signed, 1 channel. <context_class_id> Specifies format of context packets Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 148: Odi:producer:Deactivate

    <packet_size_limit> Specifies a maximum size for packets, in bytes. 0 indicates default of 262144. :ODI:PRODucer<N>:DEACtivate Command :ODI:PRODucer<N>:DEACtivate Parameters None Description This command deactivates the stream, stops any data flow, and frees resources. Example :ODI:PROD1:DEAC Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 149: Trigger Commands

    <N> An integer, 1-4. EXTernal Use the external trigger input. CHANnel<N> Use an analog input channel. Query :TRIGger:EDGE:SOURce? The query returns the current setting for the trigger source. Returned Format [:TRIGger:EDGE:SOURce] {EXTernal | CHANnel<N>}<NL> Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 150: Trigger:edge:slope

    <N> An integer, 1-4. <level> The trigger level in Volts as a floating point number. Query :TRIGger:LEVel[:CHANnel]? CHANnel<N> The query returns the current trigger level. Returned Format [:TRIGger:LEVel[:CHANnel]] <level><NL> Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 151: Trigger:holdoff

    The trigger event is associated with the time the level was crossed. Query :TRIGger:HOLDoff:TYPE? The query returns the current setting for the holdoff type. Returned Format [:TRIGger:HOLDoff:TYPE] { NORMal | ABOVe | BELow }<NL> Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 152: Test Commands

    This query returns the result of the power on self-tests. Example Query :TEST:PON? :TEST:TST? Query :TEST:TST? Parameters None Description This query is similar to *TST? but the actual test messages are returned. Example Query :TEST:TST? Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...
  • Page 153: Characteristics

    Keysight M8131A 16/32 GSa/s Digitizer User’s Guide Characteristics / 153 Performance Specification / 153 Operating Environment / 154 General Performance Specification The performance specification can be found in the Data Sheet of the M8131A at: http://www.keysight.com/find/M8131A. Operating Environment Storage Temperature −40 °C to +70 °C...
  • Page 154 Calibration interval 2 years recommended Cooling When operating the M8131A choose a location that provides at least 80 mm of clearance at rear, and at Requirements least 30 mm of clearance at each side for the AXIe chassis. Keysight M8131A 16/32 GSa/s Digitizer User’s Guide...