Trigger:level:external; Trigger:level[:Channel]; Trigger:level:ifmagnitude - Keysight Technologies M8131A User Manual

16/32 gsa/s
Hide thumbs Also See for M8131A:
Table of Contents

Advertisement

7
Remote Programming

:TRIGger:LEVel:EXTernal

Returned Format

:TRIGger:LEVel[:CHANnel]

Returned Format

:TRIGger:LEVel:IFMagnitude

176
Command
:TRIGger:LEVel:EXTernal <level>
Description
The command specifies the trigger level when the external trigger input is
used as trigger source.
<level> The trigger level in volts as a floating-point number.
Query
:TRIGger:LEVel:EXTernal?
The query returns the current trigger level.
[:TRIGger:LEVel:EXTernal] <level><NL>
Command
:TRIGger:LEVel[:CHANnel] CHANnel<N>, <level>
Description
The command specifies the trigger level when an analog input channel is
used as trigger source.
<N> An integer, 1-4.
<level> The trigger level in volts as a floating point number.
Query
:TRIGger:LEVel[:CHANnel]? CHANnel<N>
The query returns the current trigger level.
[:TRIGger:LEVel[:CHANnel]] <level><NL>
Command
:TRIGger:LEVel:IFMagnitude CHANnel<N>, <level>
Description
The command specifies the Intermediate Frequency Magnitude trigger
level when an analog input channel is used as trigger source and the data
processing mode is DDC.
A trigger event is generated, when the power P
pair after DDC processing and decimation is above the specified trigger
level in dBm according to the following formula:
P
Sample
P
Sample,dBm
2
2
2
= (i
+ q
) * yInc
/ 50 Ω
= 10 * log
(P
10
Sample
Keysight M8131A 16/32 GSa/s Digitizer User's Guide
Sample,dBm
/ 1 mW)
of an IQ sample