Channel Commands
:CHANnel<N>:DISPlay
Returned Format
:CHANnel<N>:OFFSet
Returned Format
Keysight M8131A 16/32 GSa/s Digitizer User's Guide
The CHANnel subsystem contains commands and queries for the
following:
•
Selection of channels for memory capture.
•
Configuration of voltage range and offset for digitizer modules with
single-ended inputs
•
Configuration of Digital-Down-Conversion (DDC), if the option -DDC is
installed.
•
Configuration of the correction filter.
Command
:CHANnel<N>:DISPlay {{ON | 1} | {OFF | 0}}
Description
The command selects or deselects a channel for waveform data capture to
memory. Additionally, it turns the data display of the specified channel on
or off.
<N> An integer, 1-4
Query
:CHANnel<N>:DISPlay?
The query returns the current display condition for the specified channel.
[:CHANnel<N>:DISPlay] {1 | 0}<NL>
Command
:CHANnel<N>:OFFSet <offset>
Description
The command sets the voltage value that is represented at the center of
the display for the selected channel.
<N> An integer, 1-4
<offset>
The offset, valid values are -400 mV to 400 mV.
Example
:CHAN1:OFFS 2.00000000000000E-01
Query
:CHANnel<N>: OFFSet?
The query returns the current offset value for the specified channel.
[:CHANnel<N>:OFFSet] <offset><NL>
Remote Programming
129