System Commands; System:error[:Next]; System:help:headers - Keysight Technologies M8131A User Manual

16/32 gsa/s
Hide thumbs Also See for M8131A:
Table of Contents

Advertisement

System Commands

:SYSTem:ERRor[:NEXT]?

:SYSTem:HELP:HEADers?

Keysight M8131A 16/32 GSa/s Digitizer User's Guide
Query
:SYSTem:ERRor?
Description
The query read and clear one error from the instrument's error queue.
A record of up to 30 command syntax or hardware errors can be stored in
the error queue. Errors are retrieved in first-in-first-out (FIFO) order. The
first error returned is the first error that was stored. Errors are cleared as
you read them.
If more than 30 errors have occurred, the last error stored in the queue
(the most recent error) is replaced with "Queue overflow". No additional
errors are stored until you remove errors from the queue.
If no errors have occurred when you read the error queue, the instrument
responds with 0,"No error".
The error queue is cleared by the *CLS command, when the power is
cycled, or when the Soft Front Panel is re-started.
The error queue is not cleared by a reset (*RST) command.
The error messages have the following format (the error string may contain
up to 255 characters):
error number,"Description", e.g.
-113,"Undefined header".
Example
Query
:SYST:ERR?
Query
:SYSTem:HELP:HEADers?
Description
The query returns all SCPI commands and queries and IEEE 488.2
common commands and common queries implemented by the instrument.
The response is a <DEFINITE LENGTH ARBITRARY BLOCK RESPONSE
DATA> element. The full path for every command and query is returned
separated by linefeeds.
Remote Programming
151
7

Advertisement

Table of Contents
loading

Table of Contents