Advertisement

5. Insert the signal cable through this cable gland
6. Insert the stripped conductors of the signal cable into the terminal block in accordance
with the terminal assignment.
7. Fasten the signal cable in the cable gland
of 3 Nm.
8. Insert the shielding plate
9. Close the door of the wall-mounted device and tighten the six screws
8.2.5

Connecting the Ethernet cable

Establishing a shield for the Ethernet cable
Note
Metalized cable gland
The metalized cable gland required for shielding is included in the scope of delivery.
1. Prepare the Ethernet cable:
②④
Figure 8-7
– Isolate the Ethernet cable
– Insert the Ethernet cable through the cable gland. Make sure that the exposed
– Screw the cable gland closed.
Wall-mounted device
Operating Instructions, 05/2018, A5E31930403-06
Cable gland
Contact shielding
Ethernet cable
Shielding braid
Ethernet cable shielding
over a length of approx. 1 cm.
shielding braid rests on the shield contact
.
. To do this, tighten the union nut to a torque
in the wall-mounted device again and fasten it with screws.
at the desired location so that the braid
or
Connecting the device
8.2 Electrical connections
again.
is exposed
.
93