Danger due to electrical current
The power cable to be connected at the customer location could be damaged by improper
connection to the device.
• Check whether the existing line voltage corresponds to the line voltage specified on the
• Before starting up the device, you should carry out a visual inspection and PE conductor
• Because part of the cable runs in the device, always adhere to the product safety
Missing PE/ground terminal
Danger of electrocution
If there is no ground terminal, there is a risk of electrocution.
Depending on the device version, connect the power supply as follows:
• Power plug: Ensure that the used socket has a PE/ground terminal. Check that the
• Connecting terminals: Connect the terminals according to the terminal assignment table.
Equipotential bonding
Operating the device without connection of the equipotential bonding can result in injury
and/or material damage and harm to the environment.
• Connect the device to equipotential bonding.
Wall-mounted device
Operating Instructions, 05/2018, A5E31930403-06
WARNING
nameplate of the device or in the technical specifications.
You can also find the required line voltage in the test certificate valid for your country.
test of the cable. To do this, proceed in accordance with the country-specific safety
standards, regulations or other applicable directives.
standard IEC/EN 61010-1 or its national equivalent.
WARNING
PE/ground terminal of the socket and power plug match.
Connect the PE/ground conductor first.
CAUTION
Connecting the device
8.2 Electrical connections
89