Installing / removing and connecting analyzer and option modules
7.7 Installing option modules
Performing installation
Figure 7-12
1. Inserting the terminal block:
– Insert the "Options" terminal block
– Tighten the 4 screws with a torque of 2.0 Nm.
2. Connect both terminal blocks with the existing connecting cable
3. Install the cable gland plate including both terminal blocks
nuts with a torque of 0.8 Nm.
4. Tighten the screw for the cable gland plate again with a torque of 2.0 Nm.
74
"Options" terminal block
"Standard" terminal block
Inserting and connecting terminal blocks
block
.
Connecting cable terminal blocks
above the factory-installed "Standard" terminal
Operating Instructions, 05/2018, A5E31930403-06
.