Commissioning
9.1 General safety instructions for commissioning
Hot surfaces resulting from hot sample gases and heated devices
Danger of burns resulting from surface temperatures above 70 °C (155 °F).
• Take appropriate protective measures, for example contact protection.
• Make sure that the maximum ambient temperature is not exceeded by implementing
NOTICE
Formation of condensation in the device
Condensation can be caused by a high difference in temperature between transportation
and storage and the installation location.
• Before taking the device into operation let the device adapt for several hours in the new
In the version with the order suffix -B06 (Cleaned for O
are cleaned and absolutely grease-free.
Ignition hazard
Because of the danger of ignition, all parts coming into contact with oxygen must be clean.
This means they must be free of loose parts or parts which could become loose during
operation and of foreign particles, especially oil, grease and solvents.
Observe the following when working on analyzers with specially cleaned gas path:
• No clothing which is contaminated by oil or grease may be worn.
• The installation site must be clean and dust-free.
• Only parts which are packed and marked accordingly may be assembled.
• Always wash your hands before commencing work.
• The cleaned parts must only be handled on the surfaces which do not come into contact
• New filters, new connecting hoses, and oil-free mixing equipment must always be used
104
CAUTION
protective measures. Refer to the information in the section "Technical specifications
(Page 141)".
environment.
CAUTION
with the sample gas.
to check the analyzer function.
), all parts wetted by the sample gas
2
Operating Instructions, 05/2018, A5E31930403-06
Wall-mounted device