Connection Information - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis wall-mounted device
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Connecting the device
8.1 Gas connections
8.1.2

Connection information

Note
Identifier of the gas connections
The identifiers of the gas connections are imprinted as numbers 1 to 4 on the module plate.
Note
Pipes
Connect the pipes with the gas connections using suitable clamping screws.
Use only such connecting parts (pipes, threaded joints, etc.) coming into contact with sample
gas that are resistant to the sample gas.
Note
Leaky gas connections
A gas-tight connection cannot be achieved if you install the clamping ring connections before
a sample gas restrictor or clamping screw was screwed onto the connecting socket. The gas
connection is then irreversibly leaky. Leaky gas paths lead to accumulation of the sample
gas in the device.
• Observe the information provided by the manufacturer of the clamping screws. Observe
the installation guidelines.
• Before you install the clamping ring connections, make sure that the connecting socket is
either equipped with a sample gas restrictor or with a clamping screw.
• Tighten the clamping screws in accordance with the mounting specifications of the
manufacturer using a suitable open-ended wrench. In the process, make sure that you
counter properly.
80
Operating Instructions, 05/2018, A5E31930403-06
Wall-mounted device

Advertisement

Table of Contents
loading