3. Remove the equipotential bonding cable
Note: The cable is no longer required for equipotential bonding
module
Figure 7-3
4. Loosen the eight Torx screws
5. Check the flat seal
Replace the flat seal
6. Swivel the analyzer module
7. Fasten the analyzer module
– Use eight Torx screws on the base of the enclosure
– Use two Torx screws on the rear panel with a torque of 2.5 Nm.
Wall-mounted device
Operating Instructions, 05/2018, A5E31930403-06
⑥
is installed.
Unlatching plug-in connectors at the cable equipotential bonding
NOTICE
Damage to the device through pulling the latched cable
Cables for equipotential bonding are equipped at their two ends with latching plug-in
connectors. Pulling on a latched cable can cause the plug-in connector to be torn off or
its contacts to be damaged.
• Before pulling out the plug connector, loosen the latch by pressing with your index
finger!
④
for damage.
④
Installing / removing and connecting analyzer and option modules
③
from the blank plate of the respective slot.
⑤
and remove the blank plate at the corresponding slot.
if it is damaged or porous.
⑥
forward and insert it into the wall enclosure
⑥
at the mounting location as follows:
7.4 Installation of analyzer modules
③
after the analyzer
①
⑤
with a torque of 0.8 Nm.
.
63