Siemens SIPROCESS GA700 Operating Instructions Manual page 63

Continuous gas analysis wall-mounted device
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

3. Remove the equipotential bonding cable
Note: The cable is no longer required for equipotential bonding
module
Figure 7-3
4. Loosen the eight Torx screws
5. Check the flat seal
Replace the flat seal
6. Swivel the analyzer module
7. Fasten the analyzer module
– Use eight Torx screws on the base of the enclosure
– Use two Torx screws on the rear panel with a torque of 2.5 Nm.
Wall-mounted device
Operating Instructions, 05/2018, A5E31930403-06
is installed.
Unlatching plug-in connectors at the cable equipotential bonding
NOTICE
Damage to the device through pulling the latched cable
Cables for equipotential bonding are equipped at their two ends with latching plug-in
connectors. Pulling on a latched cable can cause the plug-in connector to be torn off or
its contacts to be damaged.
• Before pulling out the plug connector, loosen the latch by pressing with your index
finger!
for damage.
Installing / removing and connecting analyzer and option modules
from the blank plate of the respective slot.
and remove the blank plate at the corresponding slot.
if it is damaged or porous.
forward and insert it into the wall enclosure
at the mounting location as follows:
7.4 Installation of analyzer modules
after the analyzer
with a torque of 0.8 Nm.
.
63