Reference Gas Monitoring; Calomat 7; Module Design - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis wall-mounted device
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Description

4.6 CALOMAT 7

4.5.3

Reference gas monitoring

Only the device versions of the ULTRAMAT 7 with a flow-type reference end are equipped
with reference gas connections. The reference gas connections are implemented as nozzles
with an outer diameter of 6 mm.
To connect the reference gas supply and return lines use only such gas-contacted
connecting parts that are suitable for the connection and the reference gas.
The flow rate of the reference gas can range from 0.1 l/min to 1.5 l/min. We recommend a
flow rate of approx. 0.5 l/min.
For devices without flow-type reference compartment, the reference cuvette is filled with
nitrogen and hermitically sealed.
4.6
CALOMAT 7
4.6.1

Module design

Figure 4-8
32
Sample gas inlet
Sample gas outlet
Not assigned
Not assigned
Design of CALOMAT 7, sample gas path with pipes
Power supply connection
Connection optional module 2.x (OM2.x)
Connection internal bus communication
Operating Instructions, 05/2018, A5E31930403-06
Wall-mounted device