Application Planning; General Safety Instructions - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis wall-mounted device
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

5.4

Application planning

5.4.1

General safety instructions

Introduction of combustible gases
Explosion hazard
Devices in standard version are not designed for use in hazardous areas.
• Do not use a standard version of the analyzer in hazardous areas.
• Supply gases with flammable components at concentrations above the lower explosion
Introduction of toxic, corrosive or flammable gases
The limited release of toxic or corrosive gases during their introduction cannot be avoided
with absolute certainty.
• Before toxic, corrosive or flammable gases are introduced, carry out a leak test for the
• Flush the device with oil-free and dry purge air or inert gas.
• Collect the emerging purging gas for environmentally-friendly disposal with a suitable
Parts in contact with sample gases unsuitable for sample gas
Danger of injury or poisoning, device damage
Hot, poisonous and corrosive sample gases can be released at the connections if the
device parts and accessories coming into contact with sample gas are unsuitable for the
sample gas.
• Use only such connecting part (pipes, threaded joints and sealing material) coming into
Wall-mounted device
Operating Instructions, 05/2018, A5E31930403-06
WARNING
limit (LEL) only in devices with pipes.
WARNING
pipe connections.
device.
WARNING
contact with sample gas that are suitable for the connection and the sample gases.
Refer to the information in the section "Technical specifications (Page 141)".
Combined operation
5.4 Application planning
41

Advertisement

Table of Contents
loading

Table of Contents