Combined operation
5.5 Plug & measure permitted activities
5.5.2

Notes on configuration change

NOTICE
Damage caused by inadmissible procedure/faulty operations
Inadmissible procedure and/or faulty operations can result in damage to the device or its
modules/electronic equipment in connection with Plug & Measure. Therefore:
• Observe the specifications regarding installation/removal of the device hardware.
• Carefully read the following information on different Plug & Measure cases.
NOTICE
Deletion of the analyzer module parameter assignment due to faulty operation during
commissioning
During commissioning after replacement of the AM electronics, a faulty operation can
prevent transfer of the parameters and render the AM electronics unusable. When you
commission the device once again after replacing the AM electronics, the following applies:
• First, confirm the message regarding the change of the configuration.
• Make sure to confirm the transfer of the parameters by pressing <ENTER>.
• Always carry out a normalization. To do this calibrate the analyzer module in the service
Note
Authorization level
To execute the actions described below, you must be logged on with at least the "Expert"
authorization level. You may have to log on again with the requested PIN.
Message panel for configuration change
When you change the hardware equipment of the device by replacing or adding modules or
components, the LUI alarm window "Device configuration changed" is displayed during re-
commissioning.
To start the data synchronization, confirm the alarm by pressing <ENTER>.
Depending on the concrete application adapt the required parameter sets in the message
panel "Transfer parameters". Information on the further procedure is available in the tables
below.
48
menu as follows:
"Main menu" > "2. Settings" > "20. Service" > "03. Factory calibrations" > "03.
Normalization".
Operating Instructions, 05/2018, A5E31930403-06