Combined operation

5.2 Operating modes

5.2
Operating modes
5.2.1

Setting note

Note
Operating mode-specific parameter assignment of the gas path in serial/parallel operation
Analyzer modules can be operated in series or in parallel.
• If you are only using one analyzer module or want to operate two analyzer modules in
series, set the value "Shared" in the "Gas path" parameter field.
• If you are operating two analyzer modules in parallel, however, select the setting
"Separate".
You can find the corresponding setting options in the menu → [2.12].
5.2.2

Serial/parallel operation

If you are using two analyzer modules in one basic device, the analyzer modules can be
operated in series or in parallel:
● Serial operation: The sample gas flows through the analyzer modules consecutively.
To do this, connect the sample gas outlet of the analyzer module filled first to the sample
gas inlet of the second analyzer module.
● Parallel operation: The sample gas flows separately through the analyzer modules.
To do this, connect each sample gas inlet individually to the sample gas in such a way
that the analyzer modules are supplied separately with sample gas.
5.2.3

Pressure operation/Suction operation

The analyzer modules can be operated in pressure operation and in suction operation:
● Pressure operation: The sample gas is fed into the analyzer module with overpressure
from the gas line or gas bottle.
● Suction operation: The sample gas is sucked out of the analyzer module by a
downstream pump.