Installing the device

6.3 Installing the wall-mounted device

Note
Loss of degree of protection in case of open enclosure
Damage to device if the enclosure is permanently open or not properly closed.
The degree of protection stipulated on the rating plate or in the technical specifications is no
longer guaranteed if the device is open or not completely closed.
• Install the device in accordance with the mounting instructions in order to ensure the
degree of protection.
• Close the enclosure after finishing the mounting or maintenance work.
Note
Measuring high oxygen concentrations
For the measurement of high O
that all parts that come in contact with the sample gas are free of oil and grease and do not
contain any particles that could react with O
We highly recommend you use specially cleaned analyzer modules with the "Cleaned for
O2" option for such measurements.
6.3
Installing the wall-mounted device
Dimension drawing
The following figure shows the dimensions of the wall-mounted device and the distance of
the fastening holes. The dimensions are specified in millimeters.
Figure 6-1
56
concentrations in excess of 25%, it is absolutely necessary
2
Drilling template and side view
.
2
Operating Instructions, 05/2018, A5E31930403-06
Wall-mounted device