Technical Data Ultramat 7 - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis wall-mounted device
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Technical specifications
13.3 Technical specifications of analyzer modules
13.3.2

Technical data ULTRAMAT 7

Table 13- 17 ULTRAMAT 7: General technical specifications
General
Information on the smallest
span
Max. power consumption
Weight (only module)
Electrical safety
Max. EMC error during opera-
tion with one and/or two
OXYMAT 7/ ULTRAMAT 7 units
Table 13- 18 ULTRAMAT 7: Measuring ranges
Measuring ranges
Number of measuring ranges
Component that
can have pa-
rameters as-
signed in the
measuring
ranges
Table 13- 19 ULTRAMAT 7: Gas inlet conditions
Gas inlet conditions
Sample gas
pressure
Pressure loss between sample gas inlet and sample
gas outlet
Sample gas flow
Sample gas temperature and humidity
148
ULTRAMAT 7
≤ 30 W
Max. 5.2 kg (standard version)
In accordance with
DIN EN 61010-1
Max. 4; parameters can be assigned freely
Component
Smallest possible span
CO
CO
2
CH
4
C
H
2
4
SO
2
NO
N
O
2
NH
3
CO/NO
CO/NO
CO 0 ... 100/NO 0 ... 300 ppm
(combination
with O7)
CO
/CO
2
Normal pressure (atmospheric
pressure compensation)
ULTRAMAT 7 high-temperature
See module nameplate
≤ 120 W
1.0% according to EN 61326-1 (2013)
0 ... 10 ppm
0 ... 5 ppm
0 ... 50 ppm
0 ... 300 ppm
0 ... 50 ppm
0 ... 100 ppm
0 ... 50 ppm
0 ... 100 ppm
0 ... 100 ppm
0 ... 100 ppm
500 to 1 500 hPa (absolute)
< 10 hPa at 1.5 l/min
18 to 90 l/h (0.3 to 1.5 l/min)
Operating Instructions, 05/2018, A5E31930403-06
Largest possible span
0 ... 100%
0 ... 100%
0 ... 100%
0 ... 100%
0 ... 100%
0 ... 3%
0 ... 100%
0 ... 100%
0 ... 10000 ppm
CO 0 ... 3000 ppm/NO
0 ... 3000 ppm
0 ... 100%
1)
Wall-mounted device

Advertisement

Table of Contents
loading

Table of Contents