Technical specifications
13.3 Technical specifications of analyzer modules
Table 13- 32 CALOMAT 7: Time response
Time response
Warm-up period at room temperature
Response characteristics
Table 13- 33 CALOMAT 7: Measuring response
Measuring response
Output signal fluctuation with device-
internal signal damping of 1 s
Detection limit
Measured value drift
Repeatability
Linearity deviation
Table 13- 34 CALOMAT 7: Influence variables
Influence variables
Ambient temperature
Sample gas pressure
Sample gas flow
152
Delayed
Standard module
display T
90
Module with integrat-
ed flame arrester
Dead time T
at 1 l/min
10
Adjustable signal damping range
< 30 min (max. accuracy after 2 h)
< 2.5 s
< 3.5 s
< 0.5 s
0 to 100 s
≤ ±0.5% of smallest measuring span according to module
nameplate
(noise bandwidth corresponds to
1% = 6σ value or 0.333% = 2σ value)
≤ 1% of the smallest span according to nameplate
≤ ± 1%/week of smallest measuring span according to
nameplate or 50 vpm H
/week
2
Whichever is greater
≤ ± 1% of the current span or 100 vpm H
≤ ± 1% of the current span or 100 vpm H
≤ 0.5% of smallest measuring span/10 K or
≤ 50 vpm H
/10 K
2
Whichever is greater
≤ ± 0.5%
of smallest measuring span/1% change in pres-
1)
sure or
≤ ± 50 vpm H
/1% change in pressure
2
Whichever is greater
≤ 0.2% of smallest measuring span/dl/min. change in flow
or
≤ 50 vpm O
/dl/min. change in flow within the permissible
2
flow range (0.3 to 1 l/min)
Whichever is greater
Operating Instructions, 05/2018, A5E31930403-06
2
2
Wall-mounted device