Language Reference
TRIGger Subsystem

Frame Trigger Period

:TRIGger[:SEQuence]:FRAMe:PERiod <time>
:TRIGger[:SEQuence]:FRAMe:PERiod?
Set the frame period that you want when using the external frame
timer trigger. If the traffic rate is changed, the value of the frame period
is initialized to the preset value.
Factory Preset
250.0 µs for Basic, cdmaOne
and *RST:
4.615383 ms, for GSM
26.666667 ms for cdma2000
10.0 ms (1 radio frame) for W-CDMA (3GPP), W-CDMA
(Trial & ARIB)
90.0 ms for iDEN
20.0 ms with rate=full for NADC, PDC
40.0 ms with rate=half for NADC, PDC
Range:
0.0 ms to 559.0 ms for Basic, cdmaOne, GSM,
cdma2000, W-CDMA (3GPP), W-CDMA (Trial & ARIB)
1.0 ms to 559.0 ms for iDEN, NADC, PDC
Default Unit:
seconds
Front Panel
Access:
Mode Setup, Trigger, Frame Timer, Period
Frame Trigger Sync Mode
:TRIGger[:SEQuence]:FRAMe:SYNCmode
EXTFront|EXTRear|OFF|RFBurst
:TRIGger[:SEQuence]:FRAMe:SYNCmode?
Selects the input port location for the external frame trigger that you
are using.
Factory Preset
and *RST:
Off
Front Panel
Access:
Mode Setup, Trigger, Frame Timer, Sync Source