Questionable Integrity Signal Negative Transition
:STATus:QUEStionable:INTegrity:SIGNal:NTRansition <number>
:STATus:QUEStionable:INTegrity:SIGNal:NTRansition?
This command determines what bits in the Questionable Integrity
Signal Condition register will set the corresponding bit in the
Questionable Integrity Signal Event register when the condition
register bit has a negative transition (1 to 0). The variable <number> is
the sum of the decimal values of the bits that you want to enable.
Factory Preset
and *RST:
Range:
Questionable Integrity Signal Positive Transition
:STATus:QUEStionable:INTegrity:SIGNal:PTRansition <number>
:STATus:QUEStionable:INTegrity:SIGNal:PTRansition?
This command determines what bits in the Questionable Integrity
Signal Condition register will set the corresponding bit in the
Questionable Integrity Signal Event register when the condition
register bit has a positive transition (0 to 1). The variable <number> is
the sum of the decimal values of the bits that you want to enable.
Factory Preset
and *RST:
Range:

Questionable Power Register

Questionable Power Condition
:STATus:QUEStionable:POWer:CONDition?
This query returns the decimal value of the sum of the bits in the
Questionable Power Condition register.
The data in this register is continuously updated and reflects the
NOTE
current conditions.
Chapter 5
0 to 32767