Calibration Comb Alignment

:CALibration:COMB
:CALibration:COMB?
Aligns the comb frequencies by measuring them relative to the internal
50 MHz reference signal.
Remarks:
A valid service password needs to be entered prior to
sending the command.
Front Panel
Access:
System, Alignments, Align Subsystem, RF
Calibration Display Detail
:CALibration:DISPlay:LEVel OFF|LOW|HIGH
:CALibration:DISPlay:LEVel?
Controls the amount of detail shown on the display while the alignment
routines are running. The routines run faster if they are off, so they do
not have to update the display.
Off - displays no trace points
Low - displays every 10th trace
High - displays every trace
Factory Preset
and *RST:
Low
Front Panel
Access:
System, Alignments, Visible Align
Align the IF Flatness
:CALibration:FLATness:IF
:CALibration:FLATness:IF?
Finds the flatness shape of the current IF setup (prefilter, mgain,
natBW). This information is then used for compensating measurements
that use FFT functionality, like the spectrum measurement. The
alignment is done frequently in the background. This same alignment
is run as part of the CAL:ALL routine.
Front Panel
Access:
Select
Setup, Auto Adjust Now